USING SIX SIGMA TO EVALUATE ANALYTICAL PERFORMANCE OF HEMATOLOGY ANALYZER

Robiul Fuadi (Clinical Pathology Department of Airlangga University/ dr Soetomo Hospital Surabaya, Indonesia)



Article Info

Publish Date
13 Apr 2019

Abstract

Background: Many medical decisions  in hospital are based on hematology examination results. We must be aware about their method performance. Sigma-metric is an excellent way to evaluate analytical performance quality. We will display the performance of our laboratory hematology analyzer, Cell Dyne Ruby, by sigma-metric analysis.Method: sigma analysis was calculated by a formula, sigma = (TEa – CV)/ Bias. Total Error Allowable  (TEa) was specified by the CLIA proficiency testing criteria. The coefficien of variant (CV) and bias data  were supplied from analyzer running three levels of control low (L), normal (N), and high (H) include following analytes: hemoglobin (Hb), Red Blood Cell count (RBC), Hematocrit (HCT), White Blood Cell count (WBC), and Platelet count (PLT).Results : Sigma value as follows Hb(L:4,33 N:6,68 H:2,62), RBC(L:3,43 N:3,84 H:3,46), HCT(L:2,52 N:1,73 H:2,27), WBC (L:7,14 N:8,44 H:6,38),and PLT (L:2,46 N:8,75 H:7,84). Average sigma value for all parameters were 4,75. Minimum sigma value for any business or manufacturing process was 3. More than 6 sigma value was world class performance.Conclusion: Hematology Analyzer Cell Dyne Ruby provides “Good” performance by sigma-metric.

Copyrights © 2019






Journal Info

Abbrev

patologi

Publisher

Subject

Biochemistry, Genetics & Molecular Biology Health Professions Medicine & Pharmacology Neuroscience

Description

Indonesian Journal of Clinical Pathology and Medical Laboratory (IJCPML) is a journal published by “Association of Clinical Pathologist” professional association. This journal displays articles in the Clinical Pathology and Medical Laboratory scope. Clinical Pathology has a couple of ...