Atom Indonesia Journal
Vol 42, No 2 (2016): August 2016

ENC Measurement for ASIC Preamp Board as a Detector Module for PET System

N. Nagara (Center for Applied Nuclear Science and Technology, National Nuclear Energy Agency, Indonesia)



Article Info

Publish Date
09 Aug 2016

Abstract

We developed a gamma ray detector with an LuAG:Pr scintillator and  an avalanche photodiode as a detector for a positron emission tomography (PET) system. Studies have been performed on the influences of gamma irradiation on application-specific integrated circuit (ASIC) preamp boards used as a detector module. As a device used in nuclear environments for substantial durations, the ASIC has to have a lifetime long enough to ensure that there will be a negligible failure rate during this period. These front-end systems must meet the requirements for standard positron emission tomography (PET) systems. Therefore, an equivalent noise charge (ENC) experiment is needed to measure the front-end system's characteristics. This study showed that minimum ENC conditions can be achieved if a shorter shaping time could be applied. Received: 22 October 2014 ; Revised: 23 Februari 2016; Accepted: 3 March 2016

Copyrights © 2016






Journal Info

Abbrev

aij

Publisher

Subject

Materials Science & Nanotechnology

Description

Exist for publishing the results of research and development in nuclear science and technology Starting for 2010 Atom Indonesia published three times a year in April, August, and December The scope of this journal covers experimental and analytical research in all areas of nuclear science and ...