Analisis Dosis Radiasi Pasien Pada Pemeriksaan CT Scan Menggunakan Aplikasi Si-INTAN

Elshaday S.B Siregar (UNUD)
Gusti Ngurah Sutapa (Udayana University)
I Wayan Balik Sudarsana (Udayana University)



Article Info

Publish Date
06 Jul 2020

Abstract

CT scan is widely used to diagnose the inside of the human body, so supervision is needed to ensure the health and safety of workers, patients and the public. One surveillance that can be done is to analyze the radiation dose of the patient on CT scan with the application of Si-INTAN. Data processing of the results of CT scan of the head, thorax and abdomen for ages 0-4 years, 5-14 years and ? 15 years using the Si-INTAN application. From the results of the data processing, the highest DRL DLP value and CTDIVOL values were obtained, for CT scan heads were 1732,8 mGycm and 31,92 mGy, the CT scan of the thorax was 2450,78 mGycm and 19,36 mGy, and for CT Abdominal Scans were 3968,85 mGycm and 19,35 mGy.

Copyrights © 2020






Journal Info

Abbrev

buletinfisika

Publisher

Subject

Earth & Planetary Sciences Electrical & Electronics Engineering Environmental Science Materials Science & Nanotechnology Physics

Description

The Journal aims to promote the theory and application in the field of physics, and to encourage a vigorous dialogue between scholars and researchers all over the world. It presents original research articles, letters as well as review articles, publishes the latest achievements and developments in ...