TELKOMNIKA (Telecommunication Computing Electronics and Control)
Vol 13, No 3: September 2015

Scanning-fluorescence Reader Based on Embedded System

Zhonglong Zhao (Xiamen University)
Xiaoping Min (Xiamen University)
Shengxiang Ge (Xiamen University)
Ningshao Xia (Xiamen University)



Article Info

Publish Date
01 Sep 2015

Abstract

To measure the concentration of C-reactive protein (CRP) in serum, a portable, scanning-fluorescence reader based on time-resolved fluoroimmunoassays was developed. The scanning-fluorescence reader integrates with the AD7707 converter, which performs at a high accuracy. The photosensitive diode acts as the photoelectric conversion device, an optical module based on optical fibers, which is able to concentrate the excitation light from an LED into a line-shape beam, was designed to sendand receive the optical signal. The device subsequently addresses waveform data using a gradient, smoothing, and binarization method. When the device measures the CRP fluorescence test strip, the results exhibited a good linearity (0.99998) and the CVs (coefficient of variation) were below 5%, which indicate high accuracy. At the same time the system is low cost and small size.