Jurnal Mechanical
Vol 7, No 2 (2016)

Pengujian Orifice Flow Meter dengan Kapasitas Aliran Rendah

Ainul Ghurri (Unknown)
S.P.G. Gunawan Tisna (Unknown)
Syamsudin Syamsudin (Unknown)



Article Info

Publish Date
16 Sep 2016

Abstract

Orifice flow meter is used in many laboratory and industrial application due because of its simple design and low cost. The present research investigated an orifice flow meter operated in relatively low flow rate. Orifice plate is a metal plate, 10 mm width provided with bevel at inlet section. The diameter ratios (β) are 0.3, 0.4, 0.5, 0.6, 0.7, and 0.8, respectively. The experiment used water with flowrate range having Reynolds number between 6827,55 and 8004,72. The fluid is flown through the orifice plate. Pressure taps are used to measure pressure at upstream and downstream of the orifice plate. The actual capacity is directly measured at an outlet reservoir; while the theoretical capacity is calculated using modified Bernoulli equation with diameter ratio (β) within the equation. The results showed that the irrecoverable pressure drop decreased with the increase of flow capacity and the diameter ratio. The discharge coefficient of the orifice flowmeter ranged between 0.3 and 1.3.

Copyrights © 2016






Journal Info

Abbrev

mech

Publisher

Subject

Engineering Industrial & Manufacturing Engineering Materials Science & Nanotechnology Mechanical Engineering

Description

Jurnal Mechanical (eISSN 2460 1888 and pISSN 2087 1880), is a peer-reviewed journal that publishes scientific articles from the disciplines of mechanical engineering, which includes the field of study (peer) material, production and manufacturing, construction and energy conversion. Articles ...