Blood Pressure Monitor Design Using MPX5050GP Pressure Sensor and Visual C# 2010 Express

Ayu Jati Puspitasari (Sekolah Tinggi Teknologi Nuklir - Badan Tenaga Nuklir Nasional)
Endarko Endarko (Institut Teknologi Sepuluh Nopember)
Iim Fatimah (Institut Teknologi Sepuluh Nopember)



Article Info

Publish Date
14 Oct 2019

Abstract

Blood pressure monitor (BPM) using oscillometric method with MPX5050GP pressure sensor has been successfully fabricated and calibrated. Oscillometric method is measurement method to observe cuff pressure oscillations. In this study, the oscillometric waveform measured by fabricated BPM used Arduino Uno which processes pulse rate, systolic and diastolic pressures. Measurement results of pulse rate, systolic and diastolic pressures will be displayed and recorded in application designed using Visual C# 2010 Express. Fabricated BPM capable of measuring pulse rate and blood pressure in the range of 40-260 mmHg with ratio of systolic and diastolic pressures were achieved 0.48 and 0.78, respectively. The system has an average error of systolic pressure, diastolic pressure, and pulse rate were -4.38 mmHg (2.72%), 0.10 mmHg (5,55%), and 4.48 bpm (6.07%).

Copyrights © 2019






Journal Info

Abbrev

jfa

Publisher

Subject

Physics

Description

JFA (Jurnal Fisika dan Aplikasinya, Abbreviation: J.Fis. dan Apl.) hanya menerbitkan artikel penelitian asli serta mengulas artikel tentang topik seputar bidang fisika (fisika teori, material, optik, instrumentasi, geofisika) dan aplikasinya. Naskah yang dikirimkan ke JFA belum pernah diterbitkan ...