Jurnal Mutiara Elektromedik
Vol. 3 No. 1 (2019): Jurnal Mutiara Elektromedik

ANALISIS KALIBRASI SUHU PADA INFANT WARMER MENGGUNAKAN METODE ECRI 415-200010301-01

Iwan Ara Rizki (Fakultas Sain dan Teknologi, Universitas Sari Mutiara Indonesia)
Hotromasari Dabukke (Fakultas Sain dan Teknologi, Universitas Sari Mutiara Indonesia)



Article Info

Publish Date
26 Jun 2019

Abstract

Infant warmer is an electromedic device which is used to provide comfort and warmth in a newborn, where the baby needs a temperature that is in accordance with the temperature in the mother’s womb,which is between 34 C - 37 C. Infant warmer calibration activities refer to the ECRI method 415-20010301. As the treshold value or the value of passing the test is +- 2 C for temperature. The results obtained from calibration are 31,89+- 0,51 C for temperature. From these results it was stated that the GIGANTE infant warmer was declared to be suitable for use.