UJI KESESUAIAN PESAWAT PANORAMIC CHEPALOMETRIC DI LABORATORIUM JURUSAN TEKNIK RADIODIAGNOSTIK DAN RADIOTERAPI, POLTEKKES KEMENKES SEMARANG (UJI MENGGUNAKAN FILM GAFCHROMIC)

Andrey Nino Kurniawan (Poltekkes Kemenkes Semarang, Indonesia)
Dwi Rochmayanti (Poltekkes Kemenkes Semarang, Indonesia)
Agustina Dwi Prastanti (Poltekkes Kemenkes Semarang, Indonesia)



Article Info

Publish Date
31 May 2023

Abstract

ABSTRACT Background: X-ray equipment used for radiographic examination according to Bapeten regulations must undergo a compliance test, both for new installation and for relocating. The Panoramic Chepalometry in the Department of Radiodiagnostics and Radiotherapy Techniques Semarang is a new device and has undergone a relocation from the initial installation. In addition, the unique condition of the Panoramic Chepalometry is direct image processing or direct radiography, so a special method is needed for testing x-ray collimation beams, namely using gafchromic films Methods: The research is experimental with a one shot case study at the Laboratory of the Department of Radiodiagnostic and Radiotherapy Techniques, Poltekkes Kemenkes Semarang. The research object is a Panoramic cephalometry device. The measuring instrument used was Raysave, and the test parameters measured were: X-ray beam, voltage accuracy, time accuracy, linearity, reproducibility, HVL, and tube container leakage. The analysis of the test results is adjusted to Bapeten regulation no. 2 of 2018 Results: Parameter test results obtained: X-ray beam is still within normal limits, voltage accuracy measurement results obtained a maximum error of 2.73% (max. limit 6%), time accuracy obtained a maximum error of 7.8% (max. limit 10 %), linearity was not measured because the mA value was fixed, reproducibility was obtained CV at a voltage of 0.0012 and radiation output of 0.0006 (CV≤ 0.05), x-ray beam quality (HVL) was obtained at 3.72 mm Al at 80 kVp (HVL limit ≥ 2.3 mm Al), and no leakage in the tube housing Conclusions: The results of the panoramic cephalometric compliance test at the Laboratory of the Department of Radiodiagnostic and Radiotherapy Techniques, Poltekkes Kemenkes Semarang, on 7 test parameters as a whole are still within normal limits according to Perka Bapeten no.2 of 2018

Copyrights © 2023






Journal Info

Abbrev

jri

Publisher

Subject

Health Professions

Description

Jurnal Radiografer Indonesia (JRI) merupakan jurnal PARI yang berisi tulisan ilmiah di bidang radiologi, antara lain Radiografi Konvensional, Radiografi Digital, CT Scan, MRI, Kedokteran Nuklir, Radioterapi dan Ilmu-ilmu terkait lainnya dalam ...