Journal of Physics and Its Applications
Vol 8, No 3 (2026): August 2026

Treatment Delivery Verification Using Thimble Chamber and IMRT Phantom

Muhammad Asif (Swat Institute of Nuclear Medicine, Oncology & Radiotherapy (SINOR) Saidu Sharif Swat, KPK, Pakistan.)
Aasia Razzaq (Institute of Nuclear Medicine and Oncology Lahore (INMOL) Lahore, Punjab, Pakistan.)
Nauman Amjid (Institute of Nuclear Medicine and Oncology Lahore (INMOL) Lahore, Punjab, Pakistan.)
Tariq Siddique (Pakistan Institute of Engineering & Applied Sciences (PIEAS) Islamabad, Pakistan.)
Habib Ahmad (Swat Institute of Nuclear Medicine, Oncology & Radiotherapy (SINOR) Saidu Sharif Swat, KPK, Pakistan.)
Shoab Shah (Swat Institute of Nuclear Medicine, Oncology & Radiotherapy (SINOR) Saidu Sharif Swat, KPK, Pakistan.)
Javaid Ali (Swat Institute of Nuclear Medicine, Oncology & Radiotherapy (SINOR) Saidu Sharif Swat, KPK, Pakistan.)



Article Info

Publish Date
31 Aug 2026

Abstract

In external beam radiotherapy, over radiation dose results in severe side-effects, but under radiation dose decreases the probability of treatment for cancer patients. The purpose of the article was to verify treatment dose delivery in radiotherapy of the cancer patients using single point dose verification method. Computed tomography (CT) simulation of intensity-modulated radiation therapy (IMRT) phantom with slot-in ion chamber was done. The IMRT phantom was set on the treatment table of LINAC (Siemens) with FC65-P Farmer type thimble chamber for radiation dose measurement. Different plans were prepared on prowess panther treatment planning system (TPS) using CT images of IMRT Phantom. The monitor units (MU) and doses were calculated for both photon algorithms i.e. Collapsed Cone Convolution Superposition (CCCS) and Fast Photon (FP) algorithms. The calculations for treatment dose verification were done for both 6MV and 15MV energies produced by LINAC (Siemens) at central axis as well as off axis beams. The percentage difference between the measured and calculated dose was less than 1% except for those plans which have greater number of posterior beams. These percentage errors were slightly greater than 1% but well within international recommended limits of ±3% as per ICRU and AAPM. The percentage error in radiotherapy treatment delivery within limit confirms accurate radiotherapy treatment dose delivery with LINAC at the institute. The couch attenuation correction factor while calculating dose on treatment planning system.

Copyrights © 2026






Journal Info

Abbrev

jpa

Publisher

Subject

Astronomy Earth & Planetary Sciences Materials Science & Nanotechnology Physics

Description

Journal of Physics and Its Applications (JPA) (e-ISSN: 2622-5956) is open access, International peer-reviewed journal that publishes high-novelty and original research papers and review papers in the field of physics including Radiation Physics, Materials, Geophysics, Theoretical Physics, ...