Electron beam radiotherapy is widely used for superficial lesion treatment; however, surface dose distribution can be modified by the application of bolus material. This study aimed to evaluate the effect of a thermoplastic elastomer (TPE) bolus on absorbed dose at specific measurement depths for electron beam energies and to assess its feasibility as an alternative bolus material. An experimental phantom-based study was conducted using a slab phantom with five repeated measurements per condition. Electron beam energies of 4, 6, 8, and 10 MeV were evaluated under open field (without bolus) and with 1 cm TPE bolus conditions. Absorbed dose was calculated according to the IAEA TRS-398 protocol. The mean absorbed dose without bolus was 2.0486 ± 0.0019 Gy (4 MeV), 2.0199 ± 0.0019 Gy (6 MeV), 2.0636 ± 0.0007 Gy (8 MeV), and 2.0440 ± 0.0011 Gy (10 MeV). With TPE bolus application, the absorbed dose changed to 1.3012 ± 0.0024 Gy, 1.8396 ± 0.0074 Gy, 1.8835 ± 0.0013 Gy, and 2.0248 ± 0.0013 Gy, respectively. Coefficient of variation for all measurements was below 0.5%, indicating high reproducibility. Wilcoxon signed-rank test demonstrated significant differences between open and bolus conditions at all energies (Z ≈ −2.03; p < 0.05). The findings indicate that thermoplastic elastomer bolus significantly modifies electron beam absorbed dose, with greater absorbed dose reduction observed at lower energies. TPE demonstrates stable dosimetric performance.
Copyrights © 2026