The discharge of raw, untreated laboratory effluent containing heavy metals poses a significant environmental risk due to the non-biodegradable nature of metal contaminants. This study validates Atomic Absorption Spectroscopy (AAS) for the determination of copper (Cu) in laboratory effluent from an educational chemistry laboratory. A calibration curve (0.5–2.5 mg/L) showed excellent linearity with a regression equation of y = 0.0771x + 0.0103 (r^2 = 0.999). Due to high Cu content, a 250-fold dilution was applied prior to analysis. The Cu concentration in the raw effluent was 427.07 mg/L. Method validation using spiked samples and replicate analysis showed satisfactory performance, with recovery of 100.64% and precision of 0.995% RSD. Matrix effects were minimized through dilution, ensuring reliable quantification. The study highlights inherently safer chemistry for accident prevention (Principle 12) and proposes Anionic Polyacrylamide (PAM) as a potential flocculation agent for future treatment. These results emphasize the importance of analytical monitoring and proper waste management in educational laboratories.Keywords: Copper, Laboratory Effluent, Method Validation, Green Chemistry, Flocculation
Copyrights © 2026