Reliable and accurate laboratory results are essential for appropriate clinical decision-making; therefore, this study assessed the variability of fasting blood glucose (FBS) and glycated hemoglobin (HbA1c) measurements across ten medical laboratories in El-Marj, Libya. A single blood sample obtained from a healthy non-diabetic male was distributed to all laboratories under standardized conditions, and the results were evaluated using mean, standard deviation (SD), and coefficient of variation (CV). The analysis showed that the mean FBS value was 77.0 mg/dL (SD = 13.2; CV = 17.1%) with a result range of 65–109 mg/dL and an acceptance rate of 75%, indicating considerable inter-laboratory variation. In contrast, HbA1c measurements demonstrated better consistency, with a mean value of 5.09% (SD = 0.31; CV = 6.1%), a result range of 4.6–5.6%, and a 100% acceptance rate. These findings indicate that laboratory variability remains a concern, particularly for FBS testing, and emphasize the importance of strengthening internal quality control systems and increasing participation in external quality assessment programs to ensure compliance with international standards and improve laboratory performance.