The sterility of a Laminar Air Flow (LAF) work zone depends on stable airflow velocity and humidity, yet both parameters are usually checked manually and periodically, so deviations are detected late. Objective: This study designs and validates a real-time Internet of Things (IoT) system for monitoring airflow velocity, relative humidity, and temperature in an LAF cabinet. Methods: The system uses an ESP32 microcontroller, a DHT22 sensor, and a TCRT5000 reflective optical sensor that reads the rotation of an indicator fan at the blower outlet to estimate air velocity. Data are shown on a 3.5-inch TFT display and sent to the Blynk platform, while blower speed is regulated by Pulse Width Modulation through an IBT-2 driver. Validation was carried out against a OneMed HTC-1 thermohygrometer (10 repetitions), a TSI VelociCalc 9565 ventilation meter at nine points of a 3 × 3 grid on the HEPA filter face with five repetitions per point, and 15 data-transmission delay measurements. Results: The mean errors were 3.52% for humidity (RMSE 2.66 %RH), 1.13% for temperature, and 3.91% for airflow velocity. The mean reference air velocity was 0.47 ± 0.02 m/s (range 0.44–0.50 m/s), with every point within 0.45 m/s ± 20% and a uniformity index of 12.8%. The mean data-transmission delay was 0.99 ± 0.05 s. Conclusion: The system meets the prototype accuracy criterion (mean error < 5%) and the near real-time monitoring criterion (delay ≤ 5 s), making it a potential low-cost alternative for remote LAF condition monitoring in educational and healthcare laboratories.