UJI KELAYAKAN PESAWAT SINAR-X TERHADAP PROYEKSI PA (POSTERO-ANTERIOR) DAN LAT (LATERAL) PADA TEKNIK PEMERIKSAAN FOTO THORAX

Kadek Miniati (Udayana University)
Gusti Ngurah Sutapa (Udayana University)
I Wayan Balik Sudarsana (Udayana University)



Article Info

Publish Date
01 Feb 2017

Abstract

Research has been conducted to determine feasibility test of the X-ray planePA and LAT projections on chest x-ray techniques. The study using a water phantom object as a substitute for patients with variations in interval distance ofthe 100-180 cm. Measurement of radiation dose X-rays performed five repetitions , measurable doses had be read on the device electrometer. Exposition factors to the PA projection using a tube voltage of 75 kV, current and time of 3,2 mAs, the irradiation field areaof (30 x 30) cm2.For the LAT projection tube voltage of 80 kV, current and time of 6,3 mAs, and the irradiation field area of (20 x 30) cm2. It the study of the radiation dose X-ray plane projection PA and LAT is optimal is below the limit value at the level of dosage guidelines BAPETEN No 08 of 2011. Obtained PA projections are below the value of 0,4 mGy while LAT projection is below the value of 1,5 mGy. The radiation dose X-rays plane using a variation of 100-180 cm distance is still below the dose limit values ??, thus meeting the objectives anssurance quality and quality control.

Copyrights © 2017






Journal Info

Abbrev

buletinfisika

Publisher

Subject

Earth & Planetary Sciences Electrical & Electronics Engineering Environmental Science Materials Science & Nanotechnology Physics

Description

The Journal aims to promote the theory and application in the field of physics, and to encourage a vigorous dialogue between scholars and researchers all over the world. It presents original research articles, letters as well as review articles, publishes the latest achievements and developments in ...