This Author published in this journals
All Journal Jurnal Instrumentasi
Claim Missing Document
Check
Articles

Found 1 Documents
Search

Development of an Automated Battery Voltage Measurement System for Preventive Maintenance Using LabVIEW and MySQL Karman, Sukarman; Paramacita, Luh
Instrumentasi Vol 50, No 1 (2026): Dalam Proses Penerbitan
Publisher : National Standardization Agency of Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31153/instrumentasi.v50i1.838

Abstract

Manual data recording for battery preventive maintenance (PM) is prone to error and inefficient. This study developed an integrated automated data acquisition system to address these issues. The implemented system utilizes an Arduino Nano with a voltage sensor and Bluetooth module for measurement and wireless transmission. A custom LabVIEW interface receives the data in real-time and provides visualization, while a MySQL database serves as a centralized repository. Testing demonstrated the sensor circuit's high reliability, showing an average measurement error of 0.50% and a Mean Absolute Error (MAE) of 0.01 V. Most significantly, the system reduced the total PM execution time by 24% by eliminating manual transcription and double-entry. This integrated solution presents a practical and cost-effective framework for modernizing PM operations in industrial settings, enhancing both data accuracy and operational efficiency.