DIFFERENCES VALUE OF PT AND APTT IN EXAMINATION OF ELECTROMECHANICAL AND PHOTO-OPTICAL METHOD Ramadhani, Beytri; Notopuro, Paulus B.
Journal of Vocational Health Studies Vol. 5 No. 1 (2021): July 2021 | JOURNAL OF VOCATIONAL HEALTH STUDIES
Publisher : Faculty of Vocational Studies, Universitas Airlangga

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.20473/jvhs.V5.I1.2021.17-21

Abstract

Background: Examination of PT and APTT in hospitals and clinical laboratories by utilizing the use of different instruments and methods. Examination of PT and APTT can be implemented through electromechanical or photo-optical techniques to detect changes in plasma turbidity. This method principle isprincipled that addition can affect the increase in plasma viscosity. Purpose: To analyze the differences in the values of PT and APTT between the electromechanical method and the photo-optical method. Method: Analytical observation, 32 plasma citrate without interference and hemolysis were examined at the Clinical Pathology Laboratory of RSU Haji Surabaya and the Ultra Medica Main Clinics Laboratory in Surabaya. The study used SPSS 24.0 program to determine whether there were  PT and APTT values with an electromechanical and photo-optical method. Result: The result of PT and APTT with the electromechanical method was significantly higher than PT and APTT with the photo-optical method on samples without interference. In hemolytic samples, the result of PT with the electromechanical method was significantly higher than the PT result with the photo-optical method. Meanwhile, the result of APTT with the electromechanical method was significantly lower than the APTT result with the photo-optical method in hemolytic samples. Conclusion: There were significant differences in PT and APTT results between electromechanical and photo-optical in samples without interference and hemolytic. It is due to the difference in the detection principle between the two methods.
DISCREPANCY INR VALUE (INTERNATIONAL NORMALIZED RATIO) IN OPTICAL AND ELECTROMECHANICAL METHOD Yunensie, Putri Adelia; Notopuro, Paulus B.
Journal of Vocational Health Studies Vol. 5 No. 1 (2021): July 2021 | JOURNAL OF VOCATIONAL HEALTH STUDIES
Publisher : Faculty of Vocational Studies, Universitas Airlangga

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.20473/jvhs.V5.I1.2021.12-16

Abstract

Background: INR examination services at laboratories in hospitals and clinics use different methods and tools therefore, results of the INR examination are different. Purpose: To determine whether there are differences in the INR value between using the photo optic method and electromechanical method. Method: This study used a citrate blood sample with a ratio of blood and anticoagulant 9: 1 and used the Sysmex CA-600 device for the photo optic method and used the Thrombostat device for the electromechanical method. The sample consisted of 32 samples, namely 10 treatment samples to be analyzed and 22 normal samples. Using SPSS 25.0 to test data that was tested for normality test and in a different test. Result: The result of INR with the electromechanical method was significantly longer than INR with the photo optic method. In the hemolytic sample, the result was significantly higher than the INR result with normal samples. Conclusion: There are significant differences between results of the INR value using photo optic and electromechanical due to the difference in the detection principle between the two methods.