Penggunaan Reagen Modifikasi pada Pemeriksaan Hemoglobin Metode Sianmethemoglobin Muhammad Ahyat Ridhoni; Rusmini Rusmini; Linda Yuliana
Journal of Educational Innovation and Public Health Vol. 4 No. 4 (2026): Oktober : Journal of Educational Innovation and Public Health
Publisher : Lembaga Pengembangan Kinerja Dosen

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.55606/innovation.v4i4.10393

Abstract

Cyanmethemoglobin method hemoglobin level measurement is still widely used as a reference method due to its good accuracy, but the relatively expensive cost of standard reagents drives the need for a more economical alternative reagent. This laboratory experimental study with a Post-test Only Control Group Design aimed to determine the difference in hemoglobin level measurement results using a modified surfactant-based reagent—formulated from 5 ml of commercial dishwashing liquid, 10 ml of analytical grade Coco Glucoside, and distilled water up to 1000 ml—compared to commercial Drabkin reagent. Measurements were performed on low, normal, and high level commercial control blood with 16 repetitions each. Data analysis included the Shapiro–Wilk normality test, paired sample t-test, and calculation of bias percentage. All data were normally distributed (p > 0.05) with mean hemoglobin levels closely approximating the Drabkin reagent across all control levels. Paired sample t-test results indicated no significant differences between the two reagents at low (p = 0.719), normal (p = 0.084), and high levels (p = 0.632), with bias percentages of −0.20%, −0.47%, and −0.06%, respectively. Consequently, the modified surfactant-based reagent produced analytical performance comparable to Drabkin reagent and has the potential to be developed as an economical alternative following further validation.