Evaluation of Platelet Count Techniques in EDTA-Induced Pseudothrombocytopenia: Sysmex XN and Mindray BC Series Saristha Buathong; Natchanon Thongrud; Kanta Durmsung; Peempol Chokchaipermpoonphol
INDONESIAN JOURNAL OF CLINICAL PATHOLOGY AND MEDICAL LABORATORY Vol. 32 No. 3 (2026)
Publisher : Indonesian Association of Clinical Pathologist and Medical laboratory

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.24293/ijcpml.v32i3.2721

Abstract

Ethylenediaminetetraacetic acid-induced pseudothrombocytopenia (EDTA-PTCP) causes falsely low platelet counts due to in vitro platelet aggregation, risking misdiagnosis. Conventional methods may be unreliable in such cases. Advanced technologies, including PLT-O (optical fluorescence-based counting) in the Sysmex XN10 and Mindray BC6200, and PLT-H (hybrid method combining optical and impedance techniques) in the Mindray BC760, aim to enhance counting accuracy. This study aimed to evaluate the performance of these techniques in both untreated and kanamycin-treated EDTA-PTCP samples. Sixty-five EDTA-PTCP samples were analyzed using Sysmex XN10, Mindray BC6200, and Mindray BC760 analyzers before and after kanamycin treatment. Manual platelet counts served as the reference standard. Statistical analysis included paired t-tests and correlation coefficients to compare analyzer performance. Platelet flagging sensitivity was assessed to evaluate the analyzers’ ability to detect abnormalities. All three analyzers demonstrated improved accuracy following kanamycin treatment, with post-treatment correlations (R-values) exceeding 0.82. The Mindray BC6200 PLT-O method showed the highest correlation both before (R = 0.774) and after treatment (R = 0.910), with minimal bias (slope = 0.982; intercept = –8.945), and outperformed Sysmex in both correlation and agreement. In terms of platelet clumping detection, the combined flagging sensitivity was highest for Mindray BC6200 and BC760 (0.800), compared to Sysmex XN10 (0.338). These findings support the use of kanamycin and advanced optical methods for improving diagnostic reliability in EDTA-PTCP cases.