Claim Missing Document
Check
Articles

Found 1 Documents
Search

Comparative Evaluation of External Calibration and Standard Addition Methods for the Determination of Lead in Instant Coffee Using Flame Atomic Absorption Spectrometry Puspitasuki, Ambar; Wibowo, Sindu; Ruwindya, Yorfan; Istiningrum, Reni Banowati
INDONESIAN JOURNAL OF CHEMICAL RESEARCH Vol. 11 No. 2 (2026): Volume 11, ISSUE 2, 2026
Publisher : Chemistry Department, Universitas Islam Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.20885/ijcr.vol11.iss2.art2

Abstract

This study compared the performance of external calibration and standard addition methods for the determination of lead (Pb) in instant coffee using flame atomic absorption spectrometry (FAAS). Accurate determination of Pb in coffee is challenging because the complex sample matrix may interfere with instrumental measurements. Therefore, both quantification methods were evaluated to determine the most suitable approach for Pb analysis. Coffee samples were prepared by dry ashing, followed by Pb determination using FAAS under identical instrumental conditions. Method performance was assessed based on calibration linearity, detection capability, and Pb concentration obtained from each quantification approach. The external calibration method produced a calibration equation of y = 0.3131x + 0.0025 with a correlation coefficient (R) of 0.9997. In contrast, the standard addition method yielded y = 0.3105x + 0.0042 with R = 0.9996, indicating excellent linearity for both methods. The Pb concentration determined by external calibration was 0.053 mg kg⁻¹, while the standard addition method gave a considerably higher value of 0.338 mg kg⁻¹. The concentration obtained by external calibration was below the method detection limit, indicating insufficient sensitivity for accurate Pb determination in the coffee matrix. In contrast, the standard addition method effectively minimized matrix interference and provided more reliable analytical results. Therefore, the standard addition method is recommended for Pb determination in instant coffee using FAAS, particularly for samples containing Pb at low concentrations.