Collimator lamp illuminance is an important parameter in the quality control of X-ray equipment because it provides a light field for determining the irradiation field before exposure. Adequate illuminance supports accurate collimation and helps ensure that the irradiation field is appropriate for the examination. According to the applicable quality control standard, collimator lamp illuminance should be at least 100 lux at a distance of 100 cm. The goals of study is evaluation of Collimator lamp illuminance on an X-Ray unit. This study used a quantitative method with a descriptive observational approach. Illuminance measurements were performed using a lux meter at a focus-film distance (FFD) of 100 cm with a field size of 25 cm × 25 cm. The light field was divided into four quadrants, namely Quadrants I, II, III, and IV. Each quadrant was measured three times. Background illuminance was also measured three times with the room lighting on and the collimator lamp off. The mean background illuminance was then used to correct the measurement results of each quadrant. The corrected illuminance values were analyzed descriptively and compared with the minimum quality control standard. The corrected mean illuminance values were 148.27 lux in Quadrant I, 175.27 lux in Quadrant II, 166.27 lux in Quadrant III, and 174.94 lux in Quadrant IV, with an overall mean of 166.19 lux. All measurement points exceeded the minimum standard of 100 lux at a distance of 100 cm. Therefore, the collimator lamp illuminance on the tested X-ray unit met the applicable quality control requirements and was considered adequate to support accurate visualization of the irradiation field and radiological diagnostic services.