Zuhdi Ismail
Pusrisbang SDM - Badan Standardisasi Nasional

Published : 1 Documents Claim Missing Document
Claim Missing Document
Check
Articles

Found 1 Documents
Search

Correction of Air Buoyancy on E0-74 Mass Measurement with Pt-Ir K-112 Standard Zuhdi Ismail; Renanta Hayu Kresiani
Journal of Physics: Theories and Applications Vol 5, No 2 (2021): Journal of Physics: Theories and Applications
Publisher : Universitas Sebelas Maret

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.20961/jphystheor-appl.v5i2.59192

Abstract

Determination of the mass value of E0 no 74 by calibrating it using the Pt-Ir standard has high complexity. This is caused by a significant difference in the density of standards. In this study, the E0 no 74 made of stainless steel has a density of 8.051.130 kg.m−3 calibrated by a mass standard of Pt-Ir which has a density of 21.552.940 kg.m−3. By Archimedes' law, the buoyant force generated by the air due to this density difference is huge. Therefore, the correction of air buoyancy is very important. The air buoyancy correction value is obtained by determining the air density using the CIPM-2007 formula. The correction that arises due to air buoyancy while weighing in the air is -0.00115 g and is greater at higher density. The implication of buoyancy correction is identified by analyzing the equivalence of the measurement result using a reference value from the calibration certificate issued by BIPM. The measurement result is equivalent if -1<En<1. The  value for the corrected mass is -0.078.