Electromagnetic interference measurement for axle counters light rapid transit railway in Indonesia Yudhistira Yudhistira; Yoppy Yoppy; Elvina Trivida; Tyas Ari Wahyu Wijanarko; Hutomo Wahyu Nugroho; Dwi Mandaris
International Journal of Electrical and Computer Engineering (IJECE) Vol 12, No 5: October 2022
Publisher : Institute of Advanced Engineering and Science

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.11591/ijece.v12i5.pp4632-4639

Abstract

The measurement and analysis of electromagnetic interference (EMI) from light rapid transit (LRT) axle counters against magnetic field interference in Indonesia has been carried out. The low-cost magnetic sensors were developed according to the British Standard (BS) EN 50592:2016. The measurement setup and magnetic field limit were based on the British Standard EN 50592:2016 and ERA/ERTMS/033281 standard. Two frequency range of the measurements, lower and higher frequencies with two different train running mode, acceleration mode and deceleration mode were applied in this research. The results in lower frequency range (10 to 100 kHz) were very close to the limit value in both acceleration and deceleration mode, especially at the 30 to 50 kHz for the y and z directions. Although there may possibly magnetic interference at low frequencies, most of the magnetic field emissions were still in acceptable range.
Conducted emission investigation of infant incubator heating control mode Khusnul Khotimah; Yoppy Yoppy; Muhammad Imam Sudrajat; Vera Permatasari; Elvina Trivida; Tyas Ari Wahyu Wijarnoko
International Journal of Electrical and Computer Engineering (IJECE) Vol 12, No 6: December 2022
Publisher : Institute of Advanced Engineering and Science

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.11591/ijece.v12i6.pp5900-5910

Abstract

This paper investigates the effect of two different heating power control systems of infant incubators on their conducted emissions. Two infant incubators which respectively employ zero-crossing control mode and phase angle control mode are observed. The research was conducted by measuring conducted electromagnetic interference (EMI) from each infant incubator's power input. Measurements are conducted both during full power condition, while the incubator's compartment temperature is far away from the temperature setpoint, and during power chopping condition, while the compartment temperature reaches the steady-state set point. Method and limit of the measurement refer to CISPR 11. It is found that conducted emission higher than the standard CISPR 11 limit occurs during power chopping on phase angle control mode. This results from the sharp rise time of voltage delivered to the heater, around 220 ns for each chopping cycle.