Intan A. S. Mu'minah
Departemen Fisika FMIPA Universitas Indonesia, Kampus UI, Depok 16424

Published : 1 Documents Claim Missing Document
Claim Missing Document
Check
Articles

Found 1 Documents
Search

VERIFIKASI DOSIMETRI PERHITUNGAN BERKAS TERBUKA PERANGKAT LUNAK IN-HOUSE TREATMENT PLANNING SYSTEM (TPS) PESAWAT TELETERAPI COBALT-60 Intan A. S. Mu'minah; La Ode Husein Z. T.; Supriyanto A. Pawiro
PROSIDING SEMINAR NASIONAL FISIKA (E-JOURNAL) Vol 4 (2015): PROSIDING SEMINAR NASIONAL FISIKA (E-JOURNAL) SNF2015
Publisher : Program Studi Pendidikan Fisika dan Program Studi Fisika Universitas Negeri Jakarta, LPPM Universitas Negeri Jakarta, HFI Jakarta, HFI

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (500.152 KB)

Abstract

Currently, cobalt-60 teletherapy units still often used for radiotherapy treatments in Indonesia. An in-house treatment planning systems (TPS) was developed by Medical Physics and Biophysics Laboratory (LFMB) Universitas Indonesia as a treatment planning software of cobalt-60 beams. Performance validation of TPS calculation is an important parameter in radiotherapy quality assurance (QA). Measurements of absorbed dose using fixed SSD treatment technique were done for 44 planning test case datasets for open beam condition with different field sizes and depths of target (isocenter). A mean dose deviations (1.810 ± 2.469) % indicate that in-house TPS calculations for open beam conditions appropriate for clinical tolerances of TPS calculation. Keywords: cobalt-60 beams, dose deviation, in-house TPS, open beams, quality assurance.