Gayuh Nurul Huda
Universitas Diponegoro

Published : 1 Documents Claim Missing Document
Claim Missing Document
Check
Articles

Found 1 Documents
Search

Portable spirometer using pressure-volume method with Bluetooth integration to Android smartphone Eko Didik Widianto; Gayuh Nurul Huda; Oky Dwi Nurhayati
International Journal of Electrical and Computer Engineering (IJECE) Vol 13, No 4: August 2023
Publisher : Institute of Advanced Engineering and Science

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.11591/ijece.v13i4.pp3977-3986

Abstract

This paper presents a study on an embedded spirometer using the low-cost MPX5100DP pressure sensor and an Arduino Uno board to measure the air exhaled flow rate and calculate force vital capacity (FVC), forced expiratory volume in 1 s (FEV1), and the FEV1/FVC ratio of human lungs volume. The exhaled air flow rate was measured from differential pressure in the sections of a mouthpiece tube using the venturi effect equation. This constructed mouthpiece and the embedded spirometer resulted in a 96.27% FVC reading accuracy with a deviation of 0.09 L and 98.05% FEV1 accuracy with a deviation of 0.05 L compared to spirometry. This spirometer integrates an HC-05 Bluetooth module for spirometry data transceiving to a smartphone for display and recording in an Android application for further chronic obstructive pulmonary disease (COPD) diagnosis.