Introduction: Dengue infection remains a major cause of hospitalization in tropical regions, particularly during the critical phase when plasma leakage, thrombocytopenia, and organ involvement increase the risk of severe complications. Hepatic dysfunction, reflected by elevated aspartate aminotransferase (AST) and alanine aminotransferase (ALT), frequently accompanies hematologic deterioration. However, the phase-specific association between platelet count and liver enzyme elevation remains insufficiently characterized. This study aimed to determine the association between platelet count and AST and ALT levels among patients in the critical phase of dengue infection. Research Methodology: An analytic observational study with a cross-sectional design was conducted at Dr. TC Hillers Regional General Hospital, East Nusa Tenggara, Indonesia, between January and December 2025. A total of 68 patients meeting the WHO criteria for critical-phase dengue were included through consecutive sampling. Platelet count, AST, and ALT levels were extracted from medical records. Data distribution was assessed using the Shapiro–Wilk test. Pearson or Spearman correlation analysis was applied as appropriate. Statistical significance was set at α = 0.05. Results: The mean platelet count was 72.4 ± 28.7 ×10³/µL. Mean AST and ALT levels were 148.6 ± 64.2 U/L and 96.3 ± 41.8 U/L, respectively. Elevated AST was observed in 95.6% of patients, and elevated ALT in 85.3%. A significant negative correlation was found between platelet count and AST (r = −0.412, p = 0.001) and between platelet count and ALT (r = −0.356, p = 0.003). Conclusion: Platelet decline is significantly associated with hepatic enzyme elevation during the critical phase of dengue infection. Integrated monitoring of platelet counts and transaminases may enhance risk stratification and support early identification of severe systemic involvement in endemic hospital settings.
Cross-sectional evaluation of two point-of-care glucose meters against the GOD-PAP reference method Muhammad Jalil Razak; Nur Ismi; Marisca Jenice Sanakya; Suprapto Suprapto
Journal Interdisciplinary Health Vol. 2 No. 3 (2026): Journal Interdisciplinary Health
Publisher : Edukasi Ilmiah Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.61099/jih.v2i3.479

Abstract

Introduction: Point-of-care testing (POCT) for blood glucose measurement is increasingly used because it provides rapid results for clinical decision-making. However, its analytical performance must be evaluated before routine implementation. This study aimed to compare the precision, accuracy, and agreement of the Easy Touch and Emtech BG Max 1 POCT devices with the Glucose Oxidase–Peroxidase (GOD-PAP) reference method for blood glucose measurement. Research Methodology: An observational analytic study with a cross-sectional design was conducted at the Buton Tengah Regional General Hospital Laboratory, Indonesia. Fifty venous blood samples were analyzed using the Easy Touch and Emtech BG Max 1 POCT devices and the GOD-PAP laboratory method as the reference standard. Precision was evaluated using the coefficient of variation (CV), accuracy was assessed through percentage bias and compliance with ISO 15197:2013, and agreement was determined using nonparametric Bland–Altman analysis. Results: The CV values were 5.54% for Easy Touch, 1.76% for Emtech BG Max 1, and 0.80% for the GOD-PAP method. The mean bias was 7.74% for Easy Touch and −5.78% for Emtech BG Max 1. Compliance with ISO 15197:2013 was achieved in 62% of Easy Touch measurements and 90% of Emtech BG Max 1 measurements. Bland–Altman analysis demonstrated good agreement between both POCT devices and the reference method. However, Emtech BG Max 1 showed narrower limits of agreement and closer agreement with GOD-PAP than Easy Touch. Conclusion: Both POCT devices demonstrated acceptable analytical agreement with the GOD-PAP method for blood glucose testing. However, Emtech BG Max 1 exhibited superior precision, lower analytical bias, and higher compliance with ISO 15197:2013, indicating greater reliability for routine clinical use. Periodic analytical performance evaluation is recommended to maintain the quality and accuracy of POCT-based blood glucose testing.