SAINS MEDIKA : JURNAL KEDOKTERAN DAN KESEHATAN
Vol 9, No 1 (2018): June 2018

Corneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients

Hilmi, Mohd Radzi (Unknown)
Mohd Kamal, Khairidzan (Unknown)
Che Azemin, Mohd Zulfaezal (Unknown)
Ithnin, Mohd Hafidz (Unknown)
Md Mustafa, Md Muziman Syah (Unknown)
Yusof@Alias, Firdaus (Unknown)
Ahmad, Norsham (Unknown)



Article Info

Publish Date
11 Nov 2018

Abstract

Introduction: Corneal curvature (CC) is an important anterior segment parameter. This study compared CC measurements conducted with two optical devices in pterygium eyes.Methods: Sixty pterygium eyes of 30 patients were enrolled in this study. CC was measured three times with the optical biometer and topography-keratometer Tomey OA-2000 (Tomey Corporation, Nagoya, Japan), then with partial optical coherence interferometry (PCI) IOL Master 500 (Carl Zeiss Meditec, Dublin, CA) and data were statistically analysed.Results: The measurements revealed in a mean CC of 43.86 ± 1.57 D with Tomey OA-2000 and 43.84 ± 1.55 D with IOL Master. Distribution of data is normal, and no significance difference in CC values was detected (P = 0.952) between the two devices. Correlation between CC measurements was highly significant (r = 0. 99; P < 0.0001). The mean difference of CC values between devices was 0.017 D and 95% limit of agreement was -0.088 to 0.12. Duration taken for measurements with the standard biometer IOL Master was longer (55.17 ± 2.24 seconds) than with Tomey OA-2000 (39.88 ± 2.38 seconds) in automatic mode. Duration of manual measurement with Tomey OA-2000 in manual mode was shorter (28.57 ± 2.71 seconds).Conclusion: In pterygium eyes, CC measured with Tomey OA-2000 and IOL Master showed similar values, and high correlation was observed between these two devices. This shows that both devices can be used interchangeably. Tomey OA-2000 is better in terms of faster to operate and has its own topography systems.

Copyrights © 2018






Journal Info

Abbrev

sainsmedika

Publisher

Subject

Health Professions Medicine & Pharmacology

Description

Sains Medika is journal of medicine and health, is a peer-reviewed and open access journal that focuses on promoting medical sciences generated from Biomedical Sciences, Public Health, Clinical Sciences, and Medical Education to integrate researches in all aspects of human health. This journal ...