Jurnal Teknokes
Vol 14 No 2 (2021): September

A Digital Pressure Meter Equipped with Pressure Leak Detection

Septian Nur Wahyu Erdyansyah (Departmen of electromedical enginering poltekkes kemenkes surabaya)
Torib Hamzah (Departmen of electromedical enginering poltekkes kemenkes surabaya)
Dyah Titisari (Department of Electromedical Engineering, Poltekkes Kemenkes Surabaya)



Article Info

Publish Date
10 Oct 2021

Abstract

A manual sphygmomanometer is an instrument used to measure blood pressure, and consists of an inflatable cuff, a mercury manometer (or aneroid gauge) and an inflation ball and gauge. To assess the condition, accuracy and safety of mercury and anaeroid sphygmomanometers in use in general practice and to pilot a scheme for sphyg- momanometer maintenance within the district. Therefore, it must be calibrated periodically. Using the MPX 5050GP sensor as a positive pressure sensor. Requires a maximum pressure of 300 mmHg. This tool is also equipped with a SD Card as external storage. The display used in this module is TFT Nextion 2.8”. After conductings measurements of the three comparisons consisting of Multifunction, DPM and mercury tensimeter to 6 times, the smallest result 0 mmHg and the largest results 251.52 mmHg. While the error in mercury tensimeter’s of leak test to module and rigel is 0.56% and 0.404%.

Copyrights © 2021






Journal Info

Abbrev

Teknokes

Publisher

Subject

Computer Science & IT Control & Systems Engineering Electrical & Electronics Engineering Engineering

Description

The JURNAL TEKNOKES is a peer-reviewed periodical scientific journal aimed at publishing research results of the medical engineering areas. The Journal is published by the Department of Medical Electronics Technology, Health Polytechnic of Surabaya, Ministry of Health Indonesia. The role of the ...