Indonesian Journal of electronics, electromedical engineering, and medical informatics
Vol 4 No 2 (2022): May

Analysis of Changes in Flow Setting Against Rise Time Using Gas Board 7500E Sensor on Bubble CPAP

Andjar Pudji (Department of Medical Electronics Engineering Technology, Poltekkes Kemenkes Surabaya)
Farid Amrinsani (Department of Medical Electronics Engineering Technology, Poltekkes Kemenkes Surabaya)
Sari Luthfiyah (Department of Medical Electronics Technology, Poltekkes Kemenkes Surabaya, Indonesia)
Lusiana Lusiana (Department of Medical Electronics Engineering Technology, Poltekkes Kemenkes Surabaya)
Shubhrojit Misra (Department of Electronics & Telecommunication Engineering, Jadavpur University, INDIA)
Nur Hasanah Ahniar (Department of Medical Electronics Technology, Poltekkes Kemenkes Jakarta II, Indonesia)
Yenda Mita Barus (Department of Electromedical Engineering Poltekkes Kemenkes Surabaya)
Lamidi Lamidi (Department of Electromedical Engineering Poltekkes Kemenkes Surabaya)



Article Info

Publish Date
28 May 2022

Abstract

Respiratory distress in neonates is one of the biggest problems encountered on a daily basis. Respiratory distress looks like rapid breathing in a newborn. CPAP Bubble is the right treatment in this case. Rise time is the time for airway pressure to reach the maximum standard value. A fast time value can make it possible to bring the breath pressure to a standard level. This study aims to analyze changes in setting flow to rise time using a gasboard sensor 7500E on a CPAP bubble device. The main advantage of this proposed method is that the results are displayed in the form of graphics, portable and there is data storage. This design uses a Gasboard 7500E sensor as a detector of oxygen concentration, a TFT LCD as a graphic display and an SD Card as data storage. The data collection is in the form of the average rise time of oxygen concentration followed by changes in its rate. Data were taken five times with three different CPAPs. The results showed that the increase in the rate affect the rise time of the CPAP oxygen concentration.

Copyrights © 2022






Journal Info

Abbrev

ijeeemi

Publisher

Subject

Computer Science & IT Control & Systems Engineering Engineering

Description

The Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics (IJEEEMI) is a peer-reviewed open-access journal. The journal invites scientists and engineers throughout the world to exchange and disseminate theoretical and practice-oriented topics which covers three (3) ...