Design of water turbidity measurement using Arduino Uno R3 on the Martapura River

Mochammad Anshori (Universitas Lambung Mangkurat)
Nurma Sari (Universitas Lambung Mangkurat)
Amar Vijai Nasrulloh (Universitas Lambung Mangkurat)



Article Info

Publish Date
01 Dec 2022

Abstract

Water is a very important daily need in human life. Water is needed for drinking, preparing food, washing and cleaning. The water used must be suitable for use with quality according to physical, biological and chemical parameters. The aim of this study is to design an instrument for water turbidity which is a one of physical parameter. Turbidity occurs due to the scattering of particles that affect the absorption of light rays in water, such as sediment. Feasible water has a turbidity value of 25 NTU. The tool for measuring water turbidity uses a turbidity sensor SKU: SEN0189 which has been integrated with the Arduino Uno R3 microcontroller and can store measurement results automatically through the Delphi 7.0 interface software in .xls format. This water turbidity measuring instrument can be applied with a smart water tool, so that the percentage value of the decrease in the level of turbidity of water that has been carried out by the filtration process can be known. The test results of the water turbidity measuring instrument showed that turbidity ≤452 NTU had a standard deviation value of ±2 NTU and turbidity ≥520 NTU had a standard deviation value of ±1. The implementation of a river water turbidity measuring instrument using a SKU:SEN0189 turbidity sensor has been carried out in Martapura river water with 3 different location points. The measurement results before filtration show that the highest turbidity value is 124 NTU at point 1, 130 NTU at point 2, and 189 NTU at point 3. In the measurement after filtration, the smallest turbidity value is 9 NTU at point 1, 9 NTU at point 2, and 13 NTU at point 3. This shows that after the filtration process of Martapura river water has been carried out with a smart water tool, Martapura river water can be used for sanitary hygiene purposes according to the water turbidity parameter with a value below the 25 NTU threshold.

Copyrights © 2022






Journal Info

Abbrev

f

Publisher

Subject

Earth & Planetary Sciences Education Materials Science & Nanotechnology Physics

Description

Jurnal Fisika FLUX is a peer-reviewed, open access journal that publishes original research articles, review articles, as well as short communication in all areas of physics including applied physics, which includes: physics instrumentation and computation, biophysics, geophysics, physics materials, ...