JHTS
Vol. 1 No. 2 (2020): Journal of Health, Technology and Science (JHTS)

Sensitivity and Specificity Tests of Hemoglobin Examination Between Hemoglobinometer and Hematology Analyzer on Pregnant Mothers

Nurul Widowaty Arbie (universitas Bina mandiri Gorontalo)
Rita Amini Warastuti (Universitas Bina Mandiri Gorontalo)
Adnan Malaha (Universitas Bina Mandiri Gorontalo)



Article Info

Publish Date
26 Dec 2020

Abstract

This study aimed to determine the sensitivity and specificity test of hemoglobin examination on pregnant women using Hemoglobinometer and Hematology Analyzer devices. The research method used a quantitative approach with the type of research being analytic observational research. The type of data used was primary data in the form of research results and secondary data in the form of data obtained from literature, books and documents. The sampling technique in this study was a purposive sampling technique, with a total sample of 73 people taken from the Kota Utara Public Health Center and Toto Kabila Regional General Hospital. The results showed that the hemoglobinometer has a sensitivity of 81% and a specificity of 100%, while the hematology analyzer has a sensitivity of 98% and a specificity of 100%.

Copyrights © 2020






Journal Info

Abbrev

JHTS

Publisher

Subject

Health Professions Medicine & Pharmacology Nursing Public Health

Description

The Journal of Health, Technology and Science (JHTS) is professionally managed to help academics, researchers and practitioners to disseminate their research results. The Journal of Health, Technology and Science (JHTS) is a peer-reviewed blind journal dedicated to the publication of quality ...