Modeling of Head Linear Accelerator (LINAC) for Study of Photon Beam Characteristics Based on GEANT4

Nurul Qomariyah (Department of Physics, Faculty of Mathematics and Natural Science, University of Mataram)
Freddy Haryanto (Department of Physics, Faculty of Mathematics and Natural Science, Institut Teknologi Bandung)
A Waris (Department of Physics, Faculty of Mathematics and Natural Science, Institut Teknologi Bandung)
R Subroto (Radiotherapy Installation, NTB Provincial General Hospital)
I WA Makmur (Radiotherapy Installation, NTB Provincial General Hospital)
Rahadi Wirawan (Department of Physics, Faculty of Mathematics and Natural Science, University of Mataram)



Article Info

Publish Date
23 Jun 2023

Abstract

A linear accelerator (LINAC) is an external radiotherapy device commonly used to treat cancer. This study aims to model the LINAC head to determine the characteristics of the photon beam generated by LINAC using a GEANT4-based Monte Carlo simulation program approach. The initial stage of the research is to build a LINAC head model. The LINAC head geometry consists of: electron source, tungsten target, flattening filter, primary collimator, X-jaw and Z-jaw, and multi-leaf collimator (MLC). The second stage is simulation data acquisition (running-beam on). In the simulation, physical interactions are in the form of empenelope, electron source particles with a pencil beam model, 0.01mm set cuts, beam on 5 x 107 history, particle energy with variations of 6 MV, 9 MV, and 12 MV, and the measuring area in a water phantom 40cm × 40cm × 40cm. This study uses a source skin distance (SSD) of 100 cm, and a radiation field area of 10cm × 10cm. The simulation results obtained a histogram of the energy spectrum distribution, percent depth dose (PDD), and beam profile (BP). The simulation results show that the energy spectrum of the third variation has the same pattern with peak energies of 0.3646 MV, 0.3837 MV, and 0.3976 MV, respectively, and the average energy of the photon beam is 0.7196 MV, 0.7745 MV, and 0.7763 MV. The value of PDD and BP gets higher along with the energy source. The simulation results show that the model can explain the differences in the photon characteristics of each energy variation.

Copyrights © 2023






Journal Info

Abbrev

jmpb

Publisher

Subject

Medicine & Pharmacology Physics

Description

Journal of Medical Physics and Biophysics is a gold open-access journal and serves as official publication media of the Indonesian Association of Physicists in Medicine, IAPM (Aliansi Fisikawan Medik Indonesia, AFISMI). JMPB publishes articles with the general concern on the application of physics ...