Journal of Medical Physics and Biophysics
Vol 10, No 1 (2023)

Analysis of Half-Value Layer on Multimeters and Manual Calculation Using Aluminum Filter

Alwali Walyatalattov Solichin (Physics Department, Faculty of Mathematics and Natural Science, Institut Teknologi Bandung)
Freddy Haryanto (Physics Department, Faculty of Mathematics and Natural Science, Institut Teknologi Bandung)



Article Info

Publish Date
21 Jun 2023

Abstract

Limited number of qualified test laboratories leads to a long queue of calibration and conformity tests. X-Ray beam quality test that utilizes the half-value layer method might be used as an option for the internal test procedures to optimize the radiation protection program by the medical physicist. RADCAL and RAYSAFE were used as the multimeters for this experiment, with the exposure setting of 53 kV, 160 mA, and 50 ms. The best result is produced by the configuration where the Al plate is placed close to the source, right below the collimator, with a 2,5% mean deviation from the multimeter’s HVL value. Meanwhile, when the Al plate is placed close to the multimeter sensor, the mean deviation can reach up to 13,5%. It can be concluded that the HVL method using an Al plate and detector can be used to measure the beam quality of the radiographic X-Ray machine. This research has no intention to replace the existing procedure but to add more insight for medical physicists’ QA/QC program. As radiation safety and protection must also be considered, not only for certifications but also actualizations.

Copyrights © 2023






Journal Info

Abbrev

jmpb

Publisher

Subject

Medicine & Pharmacology Physics

Description

Journal of Medical Physics and Biophysics is a gold open-access journal and serves as official publication media of the Indonesian Association of Physicists in Medicine, IAPM (Aliansi Fisikawan Medik Indonesia, AFISMI). JMPB publishes articles with the general concern on the application of physics ...