Makara Journal of Science
Vol. 27, No. 2

Application of Photoacoustic Imaging for Pneumonia Detection

Maqfiroh, Caesarany (Unknown)
Widyaningrum, Rini (Unknown)
Anas, Ahmad Mujtahid (Unknown)
Mitrayana, Mitrayana (Unknown)



Article Info

Publish Date
25 Jun 2023

Abstract

We used photoacoustic imaging (PAI) to visualize and compare acoustic intensity levels in pneumonia-affected and healthy chicken lungs. After histological confirmation of pneumonia, the samples were scanned and subjected to a 532-nm diode laser in a photoacoustic imaging system. The acoustic intensity level of pneumonia-affected tissue was examined and compared with that of healthy lung samples. The optimum laser frequency and duty cycle for imaging the samples were 17 kHz and 30%, respectively. The acoustic intensity levels of pneumonia-affected tissue and healthy lungs were −82.5 ± 1.8 dB and −79.9 ± 1.3 dB, respectively. We found that a simple PAI device consisting of a diode laser and condenser microphone could distinguish between pneumonia-affected and healthy lungs. Pneumonia-affected lungs produced less acoustic intensity than that healthy lungs, as supported by histological studies.