Journal of Medical Physics and Biophysics
Vol 5, No 1 (2018)

The lateral shift effect of gamma index evaluation on tomotherapy

Yosi Sudarsi Asril (Universitas Indonesia)
Nuruddin Nasution (Unknown)
Wahyu Edy Wibowo (Unknown)
Supriyanto Ardjo Pawiro (Unknown)



Article Info

Publish Date
05 Mar 2018

Abstract

Fourteen patients with different cases have been planned and treated using helical Tomotherapy HI-ART system. To investigate the lateral shift effect on gamma index, the planning verification was evaluated for a point dose and a fluence dose measurement using ion chamber and Gafchromic EBT3 film, respectively. As a result, the difference between a measured and calculated point dose was under 3% for all interest organs, which account for 1.46±0.71% (parotid), 1.97±0.56% (pelvis and abdomen), 2.27±0.24% (brain), 1.56±0.51% (nasopharynx), and 1.43±1.3% (breast). In addition, Gamma index criteria used to evaluate fluence dose was 3mm (distance-to-agreement) and 3% (dose-difference) with 90% passed rate. Overall, the results, found in this study, shows that a lateral shift of the patient table resulted in an error beyond an acceptable tolerance if the table was shifted at more than 3mm. Gamma index criteria, then, is met if the lateral shift was within 3mm.

Copyrights © 2018






Journal Info

Abbrev

jmpb

Publisher

Subject

Medicine & Pharmacology Physics

Description

Journal of Medical Physics and Biophysics is a gold open-access journal and serves as official publication media of the Indonesian Association of Physicists in Medicine, IAPM (Aliansi Fisikawan Medik Indonesia, AFISMI). JMPB publishes articles with the general concern on the application of physics ...