Concordance Level between Impulse Oscillometry and Spirometry for Pulmonary Function Test

Hapsari, Brigitta Devi Anindita (Unknown)
Melita , Melita (Unknown)
Siregar, Sihsusetyaningtyas Tiominar (Unknown)
Apriningsih, Hendrastutik (Unknown)
Adhiputri, Artrien (Unknown)



Article Info

Publish Date
30 Oct 2025

Abstract

Background: The most widely used lung function test, spirometry, requires constant coaching and more effort. An alternative modality of impulse oscillometry (IOS) was introduced, which is simpler than spirometry, requiring only basic tidal breathing and less cooperation, but with less stringent standards. Methods: This research was a diagnostic test research with medical records data of patients who underwent spirometry and oscillometry procedures within the same period. The study was conducted in Sebelas Maret University General Hospital from March to September 2024. Results: A total of 77% of examinations gave the same results, namely restriction in 32 people (30.8%) and obstruction in 48 people (46.2%). Meanwhile, as many as 23% of IOS examinations gave different results compared to spirometry examinations (discordance). Six normal results from spirometry showed the impression of obstruction in IOS. The agreement between the IOS examination and spirometry was moderately significant (P=0.0001). Conclusion: Impulse oscillometry is a useful adjunct to spirometry, which is still the gold standard, especially for patients who are incapable of performing forced maneuvers.

Copyrights © 2025






Journal Info

Abbrev

jri

Publisher

Subject

Health Professions Medicine & Pharmacology

Description

Jurnal Respirologi Indonesia (JRI) is an online and printed scientific publication of the Indonesian Society of Respirology (ISR). The journal is published thrice-monthly within a year (January, April, July and October). The journal is focused to present original article, article review, and case ...