Monitoring filterable reactive phosphate (FRP) in aquatic systems is important for eutrophication control. This study evaluated a polymer inclusion membrane (PIM)-passive sampler for FRP measurement in freshwater. The PIM was fabricated by solvent casting using poly(vinyl chloride), Aliquat 336 chloride, and 1-decanol at a 70:20:10 (w/w) ratio and characterized for morphology, elemental distribution, functional groups, wettability, and mechanical properties. The PIM exhibited a smooth, dense, non-porous structure, moderate hydrophilicity, and good mechanical stability, all of which support its suitability for carrier-mediated FRP transport. Laboratory calibration using the kinetic approach gave an FRP permeability coefficient of (6.49 ± 3.39) × 10−7 m s−1 and a sampling rate of (6.15 ± 3.20) × 10−11 m3 s−1. In a preliminary evaluation using field-collected waters exposed under controlled laboratory conditions for 24 h, the sampler yielded time-weighted average FRP concentrations of (0.91 ± 0.05) × 10−5 mol L−1 in river water, (0.34 ± 0.02) × 10−5 mol L−1 in irrigation water, and (0.30 ± 0.01) × 10−5 mol L−1 in fishpond water. The sampler performed best in river water, where matrix interference was lowest. These results indicate that the PIM passive sampler is promising for preliminary operational FRP monitoring in freshwater, although matrix-specific calibration remains necessary.
Copyrights © 2026