General Background: Dengue remains a significant public health concern, while molecular diagnosis depends on the quality of RNA in clinical specimens, particularly when sample processing is delayed. Specific Background: Prolonged storage may affect molecular detection, creating a need to assess diagnostic performance under delayed examination conditions. Knowledge Gap: Evidence comparing conventional RT-PCR and RT-LAMP for dengue detection in whole blood samples stored for extended periods remains limited within the study context. Aims: This study aimed to analyze dengue virus detection in samples stored for ≥72 hours using conventional RT-PCR and RT-LAMP and to compare the results of both methods. Results: A comparative laboratory design was conducted using 30 whole blood samples from suspected dengue patients stored at 0–4°C for 72, 96, 120, 144, 168, and 192 hours. RNA was extracted using the spin-column method. RT-PCR targeted the C–prM gene, while RT-LAMP used six primers and colorimetric detection. RT-LAMP produced positive results in 26 samples (86.7%), whereas conventional RT-PCR detected 1 sample (3.3%). Storage duration was not significantly associated with RT-LAMP results (p=0.660), while the two methods showed a significant difference in results (p<0.001). Novelty: The study compares both molecular approaches across multiple prolonged storage durations using the same clinical sample context. Implications: RT-LAMP may serve as an alternative for dengue virus detection when delayed sample handling is accompanied by declining RNA quality. Key Findings Highlights Twenty-six specimens remained positive across the examined preservation periods. Only one specimen was identified by the conventional molecular assay. No significant association occurred across the six preservation intervals. Keywords : Dengue Virus, Conventional RT-PCR, RT-LAMP, Sample Storage Duration, RNA Degradation
Copyrights © 2026