Silver(I) is highly toxic to aquatic organisms, so simple and inexpensive methods for its determination are needed. This study aimed to optimize the analytical conditions for the determination of Ag(I) by UV-Vis spectrophotometry using sodium diethyldithiocarbamate (DDTC) as the complexing agent and chloroform as the extracting solvent, and to evaluate the interference of Cu(II). The optimized parameters were the maximum wavelength, pH, shaking time and complex stability time, followed by construction of a calibration curve and an interference test in which Cu(II) (0–6 mg/L) was added to 5 mg/L Ag(I). The Ag(I)-DDTC complex showed maximum absorption at 330 nm. The optimum conditions were pH 6 and a shaking time of 15 min, and the absorbance of the extract remained relatively stable for 5–30 min after extraction (decrease ≤ 7.1%). The calibration curve was linear over 3–8 mg/L (y = 0.0504x + 0.0561; R² = 0.9927), with estimated limits of detection and quantification of 0.59 and 1.79 mg/L, respectively. Cu(II) caused a positive interference that increased with its concentration, from 7.7% at a Cu:Ag mass ratio of 0.2 to 57.5% at a ratio of 1.2. The DDTC–chloroform system is therefore suitable for the determination of Ag(I) in the absence of Cu(II), whereas samples containing Cu(II) require masking or prior separation