This paper presents a flexible wearable monopole antenna for non-invasive monitoring of human arm bone fracture healing. The antenna is fabricated on a polyethylene terephthalate (PET) substrate and is intended for integration into an arm cast for continuous monitoring during healing. With dimensions of 77 mm×53 mm×0.2 mm, the antenna operates from 1.7 to 2.6 GHz in free space and exhibits resonance changes when placed on a human arm phantom. Bone healing is assessed through changes in the antenna reflection coefficient caused by variations in the dielectric properties around the fracture site. Experimental measurements show resonance at approximately 1.84 GHz without blood, 1.88 GHz with 10 ml of blood, and 1.82 GHz with 20 ml of blood, demonstrating sensitivity to changes associated with the healing process. The simulated specific absorption rate is 0.0452 W/kg at 1 mW transmitted power, within accepted safety limits. The agreement between simulation and measurement supports the feasibility of the proposed compact, low-cost, and non-invasive monitoring approach.
Copyrights © 2026