Indonesian Journal of Chemistry
Vol 9, No 2 (2009)

DIFFERENTIAL PULSE ANODIC STRIPPING VOLTAMMETRY FOR DETERMINATION OF SOME HEAVY METALS IN URANIUM

Saryati Saryati (Technology Center for Nuclear Industrial Material, National Nuclear Energy Agency (BATAN), Jakarta)



Article Info

Publish Date
22 Jun 2010

Abstract

The direct determination of some metals impurity in uranium by using differential pulse anodic stripping voltammetry (DPASV) method at a hanging mercury drop electrode and in a carbonate buffer media was developed. It was found that the carbonate buffer show the strongest affinity for uranium and gives the best separation between the DPASV peaks of heavy metals impurities. The carbonate concentration markedly affects the oxidation and reduction the major and the minor constituents of the uranium samples. In 0.1 M carbonate buffer solution pH 10, copper, bismuth, thalium, lead, cadmium, zinc, could be determined without the removal of the uranium matrix. Recovery and relative standard deviation (RSD) of this method was in the range of 174% - 85.2% for recovery and 36.8% - 1.2% for RSD. The larger error of analytical result was obtained for Zn at low concentration. In general, the analytic results error and RSD decreased with increasing metals concentration.

Copyrights © 2009






Journal Info

Abbrev

ijc

Publisher

Subject

Chemical Engineering, Chemistry & Bioengineering Chemistry

Description

Indonesian Journal of Chemistry is an International, peer-reviewed, open access journal that publishes original research articles, review articles, as well as short communication in all areas of chemistry including applied chemistry. The journal is accredited by The Ministry of Research, Technology ...