INCU Analyzer for Infant Incubator Based on Android Application Using Bluetooth Communication to Improve Calibration Monitoring

Vijay Anant Athavale (Walchand Institute of Technology, Solapur, INDIA)
Abhilash Pati (Siksha 'O' Anusandhan Deemed to be University, Bhubaneswar, INDIA)
A K M Bellal Hossain (University of Bisha, SAUDI ARABIA)
Sari Luthfiyah (Department of Electromedical Engineering, Poltekkes Kemenkes Surabaya)
Triwiyanto Triwiyanto (Department of Medical Electronics Engineering Technology, Poltekkes Kemenkes Surabaya)



Article Info

Publish Date
16 Mar 2022

Abstract

Worldwide, over 4 million babies die within a month of birth each year. Of these, 3.9 million are in developing countries. A proportion approximately 25% of these deaths are due to complications of premature birth, most commonly inadequate thermoregulation, water loss, and neonatal jaundice. An infant incubator provides stable temperature, relative humidity, and airflow values. A periodical calibration should be applied on infant incubator to monitor the functionality. The study aims to develop a calibration device that measures temperature, humidity, airflow, and noise in the baby incubator based on an Android application with Bluetooth communication to improve the calibration monitoring process. This is to achieve a better performance of the conventional INCU analyzer. The contribution of this research is that the values of the temperature, humidity, airflow, and noise can be displayed on both devices, the INCU analyzer machine, and mobile phone; thus, the user can monitor the measurement activities wirelessly. Furthermore, the statistical calculation for all measurements can be saved on a mobile phone device. The main design consists of temperature sensor LM35, humidity sensor DHT22, airflow sensor MPX5010DP, an analog signal conditioning circuit, an Arduino Mega microcontroller, Bluetooth module HC05, and Android mobile phone. The resulting design was compared to the standard or calibrator INCU analyzer machine (Fluke Biomedical INCU II). This study found that the smallest error is -1.72%°C, -0.106 % RH, -1.727% dB, and <0.1% m/s for temperature, humidity, noise, and airflow parameters, respectively. After the evaluation process, this device can be used as an INCU analyzer to calibrate the infant incubator.

Copyrights © 2022






Journal Info

Abbrev

Teknokes

Publisher

Subject

Computer Science & IT Control & Systems Engineering Electrical & Electronics Engineering Engineering

Description

The JURNAL TEKNOKES is a peer-reviewed periodical scientific journal aimed at publishing research results of the medical engineering areas. The Journal is published by the Department of Medical Electronics Technology, Health Polytechnic of Surabaya, Ministry of Health Indonesia. The role of the ...