Prisma Sains: Jurnal Pengkajian Ilmu dan Pembelajaran Matematika dan IPA IKIP Mataram
Vol 11, No 2: April 2023

The Effect of Different Liquid on Temperature Uniformity and Stability in Microbath 7102

Waslina Rangkuti (Faculty of Mathematics and Natural Sciences, University of Sumatera Utara, Medan)
Kerista Tarigan (Faculty of Mathematics and Natural Sciences, University of Sumatera Utara, Medan)
Syahrul Humaidi (Faculty of Mathematics and Natural Sciences, University of Sumatera Utara, Medan)
Marhaposan Situmorang (Faculty of Mathematics and Natural Sciences, University of Sumatera Utara, Medan)
Erna Frida (Faculty of Mathematics and Natural Sciences, University of Sumatera Utara, Medan)
Yahya Darmawan (School of Meteorology, Climatology and Geophysics (STMKG))



Article Info

Publish Date
30 Apr 2023

Abstract

Microbath Fluke Type 7102 is used for thermometer calibration. In the calibration process, Microbath uses liquid media as heat conductor. Liquid media in Microbath during the calibration process there is a value of uniformity and temperature stability. The value of temperature uniformity and stability is an influential component in determining the value of measurement uncertainty (U95). The smaller the U95 value, the better the calibration results. This is a factor in this study to analyse the uniformity and temperature stability of liquid types of Water, Methanol and Glycol. The uniformity test method is carried out using 5 (five) point measurements, where the reference point is in the middle. The stability test method uses the measurement of one reference point. Uniformity and stability values are connected to determine the uncertainty of measurement value using the GUM (Guide to the expression of Uncertainty in Measurement) method. The analysis showed that Methanol is more homogeneous than Glycol and Water, with values of 0.0855 ºC < 0.0942 ºC < 0.1030 ºC. Water is more stable than Methanol and Glycol, with values of 0.0021 ºC < 0.0027 ºC < 0.0028 ºC. The time to stabilise Methanol is better than Water and Glycol. Methanol can be stabilised with ± 35 - 40 minutes, Water needs ± 38 - 40 minutes and Glycol needs ± 48 - 50 minutes. The relationship between uniformity and temperature stability is that the smaller the uniformity and stability values, the smaller the U95 of a calibration result. The U95 value of Methanol 0.11 ºC, Glycol 0.12 ºC and Water is 0.13 ºC.

Copyrights © 2023






Journal Info

Abbrev

prismasains

Publisher

Subject

Biochemistry, Genetics & Molecular Biology Earth & Planetary Sciences Education Electrical & Electronics Engineering Energy Engineering Environmental Science Mathematics Physics Other

Description

J-PS (Prisma Sains: Jurnal Pengkajian Ilmu dan Pembelajaran Matematika dan IPA IKIP Mataram) was published by the Faculty of Science, Engineering, and Applied Science Universitas Pendidikan Mandalika. J-PS containing scientific articles in the form of research and ...