Sains: Jurnal Kimia & Pendidikan Kimia
Vol. 12 No. 2 (2023): EDISI DESEMBER 2023

Akurasi dan Presisi Analisis Kadar Nikel (Ni) pada Sampel Nikel Laterit Menggunakan X-Ray Fluorescence Spectometry (XRF)

Yulianti Malik (Program Studi Teknik Kimia Mineral Politeknik Industri Logam Morowali)



Article Info

Publish Date
16 Feb 2024

Abstract

Testing laboratories are required to be able to produce reliable and accountable analysis results. In order to fulfill this, it is necessary to verify the method. Method verification has been carried out for accuracy and precision parameters for the analysis of nickel elements in laterite nickel ore using XRF based on the ASTM E1085-16 analysis method. The test results show that these two parameters meet the acceptance requirements, with an accuracy rate (TA) of 99.43%, and a low error rate of only around 0.57% and meet the acceptance test requirements with the Horwitz equation, namely 0.0100 ≤ 0.0156. Meanwhile, the precision (repeatability) test carried out also met the acceptance requirements which met the test requirements using the Horwitz equation with a ratio of %RSD < ½ CV Horwitz (0.4693 < 1.84005).

Copyrights © 2023






Journal Info

Abbrev

journal

Publisher

Subject

Education

Description

SAINS : Jurnal Kimia & Pendidikan Kimia Dikelola oleh Jurusan Pendidikan Kimia, FKIP Universitas Halu Oleo Kendari sejak tahun 2012 dengan Media Cetak ber-ISSN 2301-5934 dan melakukan penerbitan pertamanya pada bulan April tahun tersebut. Sejak tahun 2021 pada terbitan Volume 10 Nomor 1, mengalami ...