Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics
Vol. 4 No. 2 (2022): May

Analysis of Changes in Flow Setting Against Rise Time Using Gas Board 7500E Sensor on Bubble CPAP

Pudji, Andjar (Unknown)
Amrinsani, Farid (Unknown)
Luthfiyah, Sari (Unknown)
Lusiana, Lusiana (Unknown)
Misra, Shubhrojit (Unknown)
Ahniar, Nur Hasanah (Unknown)
Barus, Yenda Mita (Unknown)
Lamidi , Lamidi (Unknown)



Article Info

Publish Date
17 Aug 2025

Abstract

Respiratory distress syndrome (RDS) is a breathing disorder that occurs in newborns, often in premature babies born before 28 weeks of gestation. The bubble CPAP (Continuous Positive Airways Pressure) is a device used to provide positive pressure to newborns who can breathe spontaneously but are still prone to apnoea. The rise time is the time it takes for the airway pressure to reach the maximum standard value. The aim of this study is to analyze the changes in flow regulation during the rise time using a 7500E gas sensor card on a bubble CPAP probe. The method used in this study is to use the mean hijacking of the sensor to reduce the noise generated by the sensor. When analyzing the data, the researcher recorded data up to five times and calculated the mean measurement error. The research design is calibrated to confirm the correctness of the displayed values. The results of the data analysis are a mean error value of 0.88% at a setting of 30% oxygen content, 0.78% at a setting of 50%, and 0.95% at a setting of 90%. For liters per minute (LPM) at the 1 LPM and 5 LPM settings, the mean error values are 0.18 % and 0.03 % for the 10 LPM setting. From the test results with 3 bubble CPAP devices, it appears that when a high LPM setting is used, the oxygen concentration is reached faster with a mean value of ±10 seconds. The conclusion from this study shows that increasing the oxygen flow rate affects the duration of the rise in bubble CPAP oxygen concentration. The implication of this study is that this data will help add artificial intelligence to bubble CPAP to automatically determine settings by combining breathing data from patients.

Copyrights © 2022






Journal Info

Abbrev

ijeeemi

Publisher

Subject

Computer Science & IT Control & Systems Engineering Decision Sciences, Operations Research & Management Electrical & Electronics Engineering Health Professions Materials Science & Nanotechnology

Description

Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics (IJEEEMI) publishes peer-reviewed, original research and review articles in an open-access format. Accepted articles span the full extent of the Electronics, Biomedical, and Medical Informatics. IJEEEMI seeks to ...