This study developed and evaluated an Arduino-based automatic temperature data logger using a DS18B20 sensor for calorimeter practicums. The development followed the Analysis, Design, Development, Implementation, and Evaluation stages. The sensor was connected to an Arduino Uno and programmed to record temperature every 15 s, eliminating manual reading and transcription during the thermal process. Operational performance was assessed from data completeness, temperature change (ΔT), time required to reach an operational equilibrium criterion, and the range of the final three readings. In the water-mixing test, the temperature increased from 26.62 °C to 39.56 °C (ΔT = 12.94 °C), with a 0.06 °C range in the last three observations. In the metal-response test, the recorded temperature increased from 25.81 °C to 28.94 °C (ΔT = 3.13 °C), and the last three readings were identical. All planned records were stored, giving 100% data completeness. These results show that the system functioned reliably as an automatic logger and supported objective identification of thermal stabilization. The available data, however, do not establish metrological accuracy because calibration against a traceable reference and replicated uncertainty analysis were not performed. The device is therefore acceptable for automated temperature recording in introductory calorimetry, while further calibration is required before it is used for precise determination of specific heat.