Activated Partial Thromboplastin Time (APTT) is a hemostasis laboratory test used to evaluate the intrinsic and common coagulation pathways. APTT testing can be performed using either a manual method or a semi-automated instrument, such as the Coatron M1. Differences in the working principles of these methods may lead to variations in test results. This study aimed to analyze APTT test results obtained using the manual method and the semi-automated Coatron M1 method. A comparative analytical study with a cross-sectional design was conducted. Citrated plasma samples were collected from 39 third-year students of the Diploma III Medical Laboratory Technology Study Program, Poltekkes Kemenkes Kalimantan Timur. Venous blood was collected into tubes containing 3.2% sodium citrate with a blood-to-anticoagulant ratio of 9:1. APTT was measured using both methods, and the data were analyzed descriptively. Data normality was assessed using the Shapiro–Wilk test, while differences between the two methods were analyzed using the Wilcoxon Signed-Rank Test. The mean APTT value obtained using the manual method was 30 ± 2 seconds, whereas the semi-automated method (Coatron M1) produced a mean value of 29 ± 3 seconds. Statistical analysis revealed a significant difference between the two methods (p = 0.010), although all results remained within the normal reference range. Therefore, both methods can be used for APTT testing in clinical laboratories.