Claim Missing Document
Check
Articles

Found 2 Documents
Search
Journal : Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics

A Low Cost Baby Incubator Design Equipped with Vital Sign Parameters Lamidi, Lamidi; Kholiq, Abd; Ali, Muslim
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol. 3 No. 2 (2021): May
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v3i2.203

Abstract

The aim of this study was to create a compact system at a low cost in an infant incubator. Apart from stabilizing the temperature, it can also calculate the premature baby's heart rate and oxygen saturation value. The contribution of the system being built can detect the early symptoms of congenital abnormalities so that abnormalities detection becomes more practical. The temperature stability in this system uses a threshold system, where the sensor readings are compared with the set point value. Meanwhile, the SpO2 system uses a finger sensor attached to the baby's finger to detect oxygen saturation and heart rate. The signal from the sensor is processed in a signal conditioning circuit consisting of an astable circuit, filter and amplifier, LPF, and demultiplexer. Based on the results of testing and measurement, the difference produced by this tool is only 0.13 - 0.182 ° C. The biggest error in the system is 0.517%. Measurement of skin temperature was carried out on 10 respondents with an average value of 34.825 ° C. so that the difference generated from this tool is only 0.175 ° C, the difference on the Humidity parameter is 0.21%. The average BPM reading of respondent 1 was 81 beats per minute and the average SpO2 value was 97%. The average BPM reading of respondent 2 was 83 beats per minute and the average SpO2 value was 98%. You can see that the module with the Arduino control system can work automatically and is able to adjust the stability of the baby incubator according to the settings. In addition, the BPM and SpO2 systems are within tolerance so they can be used for patient measurements
Defibrillator with Monophasic and Biphasic Fibrillation Signal Wijaya, I Gusti Bagus Aditya; Wijaya, Aditya; Yulianto, Endro; Kholiq, Abd
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol. 1 No. 2 (2019): November
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v1i2.252

Abstract

Lack of knowledge about functions, working methods, and safety can pose a big risk (death). Defibrillators are electronic devices that flow the shock electrical signal (pulse) to the heart muscle to maintain myocardial depolarization that is experiencing cardiac fibrillation (ventricular fibrillation or atrial fibrillation). The purpose of this study is make a defibrillator as a learning material so that anyone can have sufficient knowledge about the working principle and function of the defibrillator tool. This defibrillator is equipped with two wave selections namely monophasic and biphasic with asynchronous and synchronous modes, the energy provided ranges from 10-50 joules with use on tools 10, 20, 30, 50 joules. The energy will be flowed to the patient by pressing the Discharge button on Paddle which equivalently is adapted to humans. Energy disposal is controlled by a driver then the relay will work alternately from the first condensator then the second condensator to Paddle. Energy testing is done by 10 times using a defibrillator analyzer in a predetermined test point, the results of the monofacic wave measurement obtained by the error value of each setting with the smallest value of 0.0024% and the largest error value of 0.023%. To produce energy (joule) that appropriate capacitor is needed with real value. Bifasic waves cannot be tested because the defibrillator analyzer measuring instrument cannot detect the energy value (joule) capacitor that has been delivered.