Claim Missing Document
Check
Articles

Found 15 Documents
Search

Analysis Of Baby Incubator Humidity Based PID with Kangaroo Mode Singgih Yudha Setiawan; Dwi Herry Andayani; Andjar Pudji; Liliek Soetjiatie; Alievya Brillianty Anugrah Kusuma
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 4 No 1 (2022): January
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v4i1.6

Abstract

The main factor that affects the parameters of the workings of humidity in the baby incubator is the sensor used to read the humidity in the room. The purpose of this study is to analyze the humidity sensor in the baby incubator using 2 humidity sensors (DHT11 and DHT22) in the different location. The manufacture of this device used an after-only design, with a comparison device of INCU Analyzer, DHT11, and DHT22 sensors. Based on the measurement, DHT11 produced a value of 46%, while DHT22 produced a value of 55.45% with BPFK standards of 50%-70%. Based on the results of measurements using the INCU Analyzer, the average error value for DHT11 is 16.05%, while DHT22 is 3.47%. Therefore, the results showed that the DHT22 sensor was more accurate to be used in baby incubators because the measurement results were under BPFK standards and produced a low error value. This can be further implemented in a baby incubator making to improve the health and safety of the babies
Infant Warmer with Digital Scales for Auto Adjustment PID Control Parameters Anita Miftahul Maghfiroh; Farid Amrinsani; Singgih Yudha Setiawan; Rafi Maulana Firmansyah; Shubhrojit Misra
Jurnal Teknokes Vol 15 No 2 (2022): June
Publisher : Jurusan Teknik Elektromedik, POLTEKKES KEMENKES Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jteknokes.v15i2.246

Abstract

Babies need temperatures that match the temperature of the mother's womb, which is between 35°C – 37°C. The latest research on infant warmer device used fuzzy method as a system for controlling temperature in infant warmers. The problem raised in the previous research is that the temperature was not evenly distributed throughout the bed at each predetermined temperature setting. When it reached the setting temperature, the warmer continued to turn on so that the bed got hotter. Therefore, the purpose of the current research is to make an infant warmer device equipped with digital scales with a temperature setting of 350C- 370C using PID control to stabilize the temperature and ensure that the heat is evenly distributed on the bed. In addition, skin temperature is also added, allowing the nurses know at which level of patient's body temperature is when observations should be made. The infant warmer in this module used an arduino microcontroller which is displayed in 7 segments, the skin sensor used is the DS18B20 temperature sensor to read the skin temperature, while the infant warmer temperature sensor used is LM35 as a PID control system. The results of the current research in making the device module were compared with the measurement results of the comparator. It was revealed that current research has obtained smallest error of 0% in temperature setting of 350C. For the comparison with the incu analyzer, the smallest error was obtained at the temperature setting of 370C with an error value of 0% on the T5 measurement. Meanwhile, the difference in skin temperature against the thermometer is 0.10C. The results showed that the temperature distributed on the module had different error values. Hence, this research can be implemented on the PID control of infant warmer system to improve the performance of infant temperature stability.
Analysis of Temperature Stability and Accuracy on the Design of Thermometer Calibrator Based on Fuzzy Logic And On/Off Control Yunik Pujiastuti; Andjar Pudji; Singgih Yudha Setiawan; Farid Amrinsani; Khongdet Phasinam
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 4 No 3 (2022): July
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v4i3.244

Abstract

A thermometer is a medical device used to measure body temperature. To maintain the accuracy of the thermometer measurement results, periodic calibration is required. Calibration is an activity to determine the conventional correctness of the indicator values of measuring instruments and measuring materials by comparing them with measurement standards that can be traced to national and international standards for units of measure and/or international and certified reference materials. Based on the results of the identification of chronological problems that have been observed, a body thermometer that measures body temperature is needed so and a calibrator is needed to maintain the accuracy of the thermometer. The purpose of this study was to analyze the Temperature Stability and Accuracy of the Body Thermometer Calibrator Based on on-Off Control and Fuzzy Logic Control. The contribution of this research to this tool will use the development of a fuzzy logic control method to produce temperature stability in the Body Thermometer Calibrator (Digital). The method used in this study used fuzzy control and on-off control. The results of this study from the suitability test obtained a maximum error of 0.2% in the fuzzy control and 0.6% in the On-Off control. The average rise time difference for the two controls was 13.53 Seconds. The average settling time difference is 130.46 seconds. The results of this study can be concluded that the Fuzzy System is better than the On / Off system so the Fuzzy system is more suitable for thermometer calibration media.
Comparative Analysis of PID and Fuzzy Temperature Control System on Infant Warmer Abdul Majid; Endang Dian Setioningsih; Abd Kholiq; Singgih Yudha Setiawan; Anilkumar Suthar
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 4 No 4 (2022): October
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v4i4.257

Abstract

Infant Warmer is a life support equipment that is used to provide heat relief for normal and premature babies who are unable to maintain their own body temperature when in a new environment. The purpose of this research is to design an infant warmer using the DS18B20 sensor to analyze the efficiency and effectiveness between PID and fuzzy temperature control. This study used temperature settings of 34 °C, 35 °C, and 36 °C. The tool used for reference from standard measurements uses a standard infant warmer. When using the PID control used in the microcontroller detects the set temperature difference value with the actual temperature, the difference in value will be input to the PID control. The error value becomes PWM input on the PID control, the PID control will process the error value and determine the output value for the heater, the temperature sensor will read the actual temperature value which will be compared again with the temperature setting, the PID control will continue to process the error value and determine the PWM output value. until the actual temperature is equal to the setting temperature or the error value is zero
Nine Channel Data Logger to Measure Temperature Distribution on Dry Stelizer based on Andoid system Anita Miftahul Maghfiroh; Singgih Yudha Setiawan; Bayu Trisono Aji Pambudi; Steyve Nyatte
Jurnal Teknokes Vol 15 No 3 (2022): September
Publisher : Jurusan Teknik Elektromedik, POLTEKKES KEMENKES Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/teknokes.v15i3.306

Abstract

A temperature measurement in sterilization is needed to find out whether the temperature setting has been reached, because if the tool is operated continuously it will have an impact on the performance of the tool. Data logger is a tool used to record time and temperature by recording. The purpose of this research is to develop technological advances with remote or automated systems that can monitor changes in temperature rise and fall. This study uses 9 types of K-type thermocouple sensors as temperature gauges which are placed at 9 specified points. A thermocouple is connected to the MAX6675 module for conversion which initially detects temperature into digital form data. Data processing uses the Arduino Mega 2560 system and the Arduino programming software processor. HC-05 is used as a data transmission of measured results that have been read where the results are displayed on Android using the Blynk application, the data sent will be in the form of Excel. This tool uses a temperature comparison from the MEMMERT UN 55 incubator in the microbiology laboratory. The error value contained in the 100°C temperature adjustment is 2.6% at a temperature of 1, the smallest error at temperatures 7 and 8 is 0.2% due to the location of the sensor far from the heater. The error value at a temperature of 150°C is 1.7%, the smallest error at temperature 5 is 0.3%, 0.5%. The error value at a temperature of 200°C is 1.8%, the smallest error at temperature 0.5%. This research is expected to make it easier for users to simultaneously monitor temperature and simplify data processing to obtain an accuracy error value in the unit under test (UUT).
Utilization of High-Power Leds as Non-Invasive KV Meter Detectors in Collimation of Lighting Moch BAgus Fatihul Ihsan; Moch Prastawa Assalim Tetra Putra; Lamidi Lamidi; I Dewa Gde Hariwisana; Tribowo Indrato; Singgih Yudha Setiawan; Mansour Asghari
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 4 No 3 (2022): August
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v4i3.239

Abstract

X-ray radiation is used to diagnose the human body. Two parameters are often used as settings when using this machine. The first is the KV value, and the second is the mA value. If an error occurs in the kV setting, it will cause inappropriate image quality, so that it will provide inaccurate information in patient examination. Likewise with the presence of excessive doses to the patient's body. To ensure that the KV value that comes out is in accordance with the settings on the machine consul, invasive and non-invasive measurements can be carried out. Non-invasive is becoming an easy standard to do. Several types of equipment on the market and research results have been widely used for this non-invasive activity. The problem is that currently the existing tools still use detectors at an expensive price. The purpose of this study was to design a low-cost non-invasive x-ray KVmeter detector using an LED detector and test the ability of the detector at each point of collimation.. The method used in this study is to stump the detectors placed at 4 ends of the collimation 20 cm apart. The data is taken by doing x-ray exposure at a distance of 60 cm. The module measurements were carried out under 80 mA exposure conditions for 1 second and a collimation area of ​​20 x 20 cm. X-ray exposure settings were performed at 40kV, 50kV, 60kV, and 70kV settings. The module measurement results are compared with the x-ray machine setting values. From the comparison results, the smallest error rate on Sensor 2 is 0.83% while the highest error is on S5 of 26.43%. The results can be concluded that the LED phosphor can capture x-rays, but the detector is weak due to interference from ambient light. The results obtained from the detector itself are still less stable and linear. In future research, stability and linearity will be built using a mechanical design that reduces ambient light interference.
Enhancing Laboratory Efficiency: A Community Service Program for Centrifuge Maintenance at Randegansari Husada Hospital Setioningsih, Endang Dian; Sumber; Setiawan, Singgih Yudha; Lusiana
Frontiers in Community Service and Empowerment Vol. 4 No. 1 (2025): March
Publisher : Forum Ilmiah Teknologi dan Ilmu Kesehatan (FORITIKES)

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ficse.v4i1.90

Abstract

The community service program conducted by Poltekkes Kemenkes Surabaya aimed to address the problems associated with equipment maintenance at the Randegansari Husada Hospital, focusing on its centrifuge equipment in the laboratory. The program included counseling and provided practical assistance to the hospital staff, preparing them with the knowledge and skills for the proper maintenance of equipment. Activities covered recording the specification data of the centrifuge units, damage detection, performance improvement using a preventive maintenance program, and performance testing using a digital tachometer. The calibration process also proved a centrifuge relative error of 1.37% at 3000 RPM within the acceptable error limit of 10%. Verification in time, at 300 seconds, resulted in a relative error of 4.89%, also within acceptable limits. Such efforts go a step further to ensure that the centrifuge works well in contributing toward better diagnostic accuracies. The program took a portion of this staffing challenge when it embarked on the training of a medical laboratory technician, insisting that improved and sustained equipment reliability meant improved health service delivery. These would then increase at the hospital to limit malfunctioning of equipment for overall quality care
ANALYSIS OF TEMPERATURE STABILITY THERMOGUN CALIBRATOR WITH FUZZY-PID CONTROL setyawanto, nanda dwi; Wisana, I Dewa Gede Hari; Setiawan, Singgih Yudha
Jurnal Teknokes Vol. 18 No. 1 (2025): March
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

Body temperature is an important parameter in medical diagnosis. Thermogun, a non-contact temperature measuring device based on infrared technology, needs to be calibrated periodically to maintain measurement accuracy. This study aims to design a Thermogun calibrator with FUZZY-PID control that uses water as a temperature stabilizer. This system is designed to maintain temperatures in the range of 32°C to 45°C. Testing of the device was carried out using the DS18B20 sensor, as well as the Arduino Mega 2560 microcontroller as the main controller, heater plate for heating and using the Nextion LCD for its display. The test results showed that the system was able to increase the speed of temperature increase with a low level of measurement error (error). From the results of the module data analysis on the standard thermometer, the temperature error using the standard thermometer was 0% at all settings between 32°C-45°C. from the module to 3 Thermoguns using FUZZY-PID control In the experiment conducted, it was found that thermogun 3 had a good average reading of all temperatures that had been measured on the module with an error between the largest 0.625% at various temperature settings. Then on thermogun 1 & 2 had a fairly large error of up to 3.15%. From the measurement results, it can be concluded that FUZZY-PID control can be used as a temperature stability control on a thermometer calibrator. This study succeeded in achieving the desired temperature with accurate accuracy and the tool functioned properly.
Embedded Digital Filter System on Microcontroller to Reduce Noise in Single Lead Electrocardiography Design Setiawan, Singgih Yudha; Miftahul Maghfiroh , Anita
Jurnal Teknokes Vol. 18 No. 1 (2025): March
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

The background of this research focuses on improving the quality of ECG signals which are often disturbed by noise, which can affect the accuracy of cardiac diagnosis. The main objective of the research is to show the effectiveness of the 6th order IIR (Infinite Impulse Response) filter in reducing noise in the ECG signal and increasing the Signal-to-Noise Ratio (SNR) value. This research contributes by applying advanced filtering methods to embedded systems and analyzing the results using the Fast Fourier Transform (FFT) to transmit frequency spectrum signals. The results showed that the use of a 6th order IIR filter resulted in an increase in SNR of 15.97 dB, indicating a significant improvement in the cleanliness of the ECG signal after the filtering process. FFT confirmed that this filter effectively reduced noise and clarified component frequency signals, supporting the success of the method used. The conclusion of this study is that the 6th order IIR filter is very effective in improving ECG signal quality by reducing noise and increasing diagnostic accuracy. Implications of these results include the potential use of this technology in medical devices for more accurate cardiac monitoring, as well as the need for further research to explore other screening methods and validation in various clinical settings. This research opens up opportunities for the development of better filter technology and integration in medical systems to improve patient care outcomes.
Wireless Blood Pressure Monitor with Android Integration: Tracking Systolic and Diastolic Parameters Sintiyah, Elisa Ayu; Sumber, Sumber; Yudha Setiawan, Singgih
Jurnal Teknokes Vol. 17 No. 2 (2024): June
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

Blood pressure measurement plays a crucial role in detecting underlying diseases in the human body. It enables the identification of conditions like heart failure, kidney failure, liver damage, and stroke, underscoring the importance of regular measurements. To facilitate independent and routine blood pressure monitoring, the development of an automatic blood pressure measuring device is essential. This research aims to design and fabricate a digital sphygmomanometer that can transmit measurements to a smartphone through the Blynk application. The blood pressure measurement is conducted using the MPX5050GP pressure sensor as the pressure detector. The device is programmed using the Esp32 microcontroller and incorporates an LCD screen to display the measurement results. The study involved measuring six participants, with each individual's blood pressure recorded ten times. The obtained measurements were then compared to those of the Omron HEM-7120 digital sphygmomanometer. The results revealed a discrepancy of ±9 mmHg in systolic values and ±7 mmHg in diastolic values between the two devices. The smallest systolic error observed was 0.4%, while the largest error reached 3%. Similarly, the smallest diastolic error was 2%, with the largest error recorded at 4.8%. The measurement errors, particularly in diastolic pressure, were influenced by the participants' fatigue, as the repeated measurements on the same arm led to slight arm movements during the process. The study demonstrated the successful transmission of measurement results to a smartphone, affirming the efficacy of the Blynk application. Additionally, the MPX5050GP sensor proved effective indetecting blood pressure. These findings highlight the potential of the developed digital sphygmomanometer as a reliable tool for blood pressure monitoring, promoting self-care and early detection of health issues.