Design a Low-Cost Digital Pressure Meter Equipped with Temperature and Humidity Parameters Bedjo Utomo; I Dewa Gede Hariwisana; Shubhrojit Misra
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 3 No 2 (2021): May
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v3i2.4

Abstract

Calibration is a technical activity which consists of determining one or more properties and characteristics of a product, process or service according to a predetermined special procedure. The purpose of calibration is to ensure measurement results comply with national and international standards. The purpose of this study is to design two mode digital pressure meter (DPM) device equipped with a thermo-hygrometer and pressure in which the design is completed with a selection mode to determine the positive and negative pressure (vacuum) using MPX 5050GP sensor as a positive pressure sensor. In this design DHT 22 sensors is used to measure the humidity and temperature. To test the leak test this device is also equipped with timer. This design uses a 2.4 inch Nextion TFT LCD screen to display data. Data analysis was performed by comparing modul with standard tools. In the measurement process, Mercury tensimeter was carried out 6 times the data and the smallest results were 0 mmHg on the module and 0 mmHg on the standard tool and the largest was 298.0 mmHg on the module and 300 mmHg on the standard tool. Data were collected in a room with a temperature of 31̊C and humidity of 87%. Finally, this design is applicable for daily used for electromedical engineer to calibrate the sphygmomanometer in the hospitals
Analysis of The Accuracy of Temperature Sensors at The Calibrator Incubator Laboratory are equipped with data storage base on Internet of Thing Candra Prastyadi; Bambang Guruh Irianto; Her Gumiwang Ariswati; Dyah Titisari; Steyve Nyatte; Shubhrojit Misra
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 4 No 4 (2022): November
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v4i3.241

Abstract

A laboratory incubator is a tool used to incubate or incubate a breed. Incubators provide optimum temperature conditions for microorganisms to grow. The incubator has a temperature regulator so that the temperature can be adjusted according to the breed to be incarnated. Incubators utilize hot-dry like ovens. The purpose of this study is to conduct testing and analyze the accuracy of thermocouple sensors with incubator media in laboratory incubator calibrator tools. The contribution of the research is to know the level of accuracy of the sustainable sensor for sensing the temperature in the lab incubator. The main Design consists of 8 MAX 6675 Standards, 8 Thermocouple type K, Arduino Mega, and SD Card Standards. The temperature not in the incubator device is measured by a Type K thermocouple sensor. Thermocouple sensor numbers 8 channels that measure the temperature at each incubator camber point. The temperature will be stored on the SD card to analyze the data and the data can be processed into the form of a graphic. Benchmarking is done using a data logger temperature tool. This is done to make the Design results are under the standards of the Standard. After comparing with the Standard get the largest error value is 3.98%, at channel T6 temperature 35 °C with ordinary incubator media and the smallest error in ordinary incubator media point T6 temperature 37 ° C which is 0.06 % and in fan incubator temperature 35 C has the largest error which is 2.98 % and the smallest error 0.86%. The conclusion of this study is that the design can work well in measuring the temperature of the incubator, as well as the system for storing readings using the SD card Design and sending data using the internet network can work well.
Waterbath Temperature Control System with Fuzzy Logic Annastadia Afifah; Levana Forra Wakidi; Her Gumiwang Ariswati; Dyah Titisari; Shubhrojit Misra
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 5 No 2 (2023): May
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v5i2.277

Abstract

Unstable temperature or being outside the control temperature of a sample will cause a change in the quality of the sample itself. The purpose of this study was to create a temperature control system on waterbath with fuzzy logic using 7 labels. Used the DS18B20 sensor as a temperature sensor, a processor in the form of an UNO arduino, a thermostat as part of safety control, and displayed on a 16x4 LCD. Temperature selection between 30°C-60C. Design research using pre-experimental methods with one type group of post-testing design research, by comparing modules with comparison tools in the form of digital thermometers. The results of the research in the manufacture of waterbath tools were conducted compared to the results of measurements in the room with a digital thermometer. Obtained the highest Error value of 0.91% at 35 °C and the lowest error of 0.049% at 60 °C. While the error value based on the setting temperature obtained the highest error value at the temperature setting of 30°C of 1.38% and the lowest error at the temperature setting of 60 °C of 0.05%. The average time required to reach the shortest setting temperature in the temperature range is 27°C-30°C for 193 seconds, and the longest time in the temperature range is 27°C-60°C for 2257 seconds. the results showed that the fuzzy method is better compared to conventional methods. The results of this study can be implemented for waterbath temperature control system to get more stable results in maintaining setting temperature.