Design and Build a Ventilator Tester with a Peak Inspiratory Flow Waveform Display as Validation using the F1031V Sensor Rusdi Pratiwo Hadi; Her Gumiwang Ariswati; Dyah Titisari; Syaifudin Syaifudin; Sari Luthfiyah; Bedjo Utomo; Liliek Soetjiatie
Jurnal Teknokes Vol 15 No 4 (2022): December
Publisher : Jurusan Teknik Elektromedik, POLTEKKES KEMENKES Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/teknokes.v15i4.472

Abstract

The ventilator is a supporter of respiratory needs which is very important for the patient so that there are several parameters that must be monitored specifically, such as the measurement of pressure and flow rate used in the ventilator system, the accuracy of which must be in accordance with the accuracy of the respirator. One of the important parameters to monitor is PIF (Peak Inspiratory Flow) which is the peak inspiratory flow rate given through the ventilator. PIF that is too high or too low can cause adverse effects on the patient. PIF monitoring can be seen through the PIF value and waveform on the PIF. Monitoring the waveform of the PIF will be very useful to improve the results of using the ventilator. The purpose of this research is to get the accuracy and precision of the sensor to display the waveform of the ventilator output. The procedure carried out is to use the F1031V sensor to detect the flow generated by the ventilator and then detect the PIF value and PIF waveform. From this research, the measurement of accuracy and precision of the F1031V sensor to detect PIF and generate a waveform graph is said to be good. This is because the highest error value is ±2.04% at the 20 LPM setting. While the value of the largest standard deviation at the 30 LPM setting is 1.517 and the greatest uncertainty value at the 30 LPM setting is 0.061. Then, the largest correction value is found in the setting of 20 LPM and 30 LPM, namely 0.4. PIF monitoring is carried out to maximize patient care and reduce the breakdown time on the ventilator.
Analysis of The Accuracy of Temperature Sensors at The Calibrator Incubator Laboratory are equipped with data storage base on Internet of Thing Candra Prastyadi; Bambang Guruh Irianto; Her Gumiwang Ariswati; Dyah Titisari; Steyve Nyatte; Shubhrojit Misra
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 4 No 4 (2022): November
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v4i3.241

Abstract

A laboratory incubator is a tool used to incubate or incubate a breed. Incubators provide optimum temperature conditions for microorganisms to grow. The incubator has a temperature regulator so that the temperature can be adjusted according to the breed to be incarnated. Incubators utilize hot-dry like ovens. The purpose of this study is to conduct testing and analyze the accuracy of thermocouple sensors with incubator media in laboratory incubator calibrator tools. The contribution of the research is to know the level of accuracy of the sustainable sensor for sensing the temperature in the lab incubator. The main Design consists of 8 MAX 6675 Standards, 8 Thermocouple type K, Arduino Mega, and SD Card Standards. The temperature not in the incubator device is measured by a Type K thermocouple sensor. Thermocouple sensor numbers 8 channels that measure the temperature at each incubator camber point. The temperature will be stored on the SD card to analyze the data and the data can be processed into the form of a graphic. Benchmarking is done using a data logger temperature tool. This is done to make the Design results are under the standards of the Standard. After comparing with the Standard get the largest error value is 3.98%, at channel T6 temperature 35 °C with ordinary incubator media and the smallest error in ordinary incubator media point T6 temperature 37 ° C which is 0.06 % and in fan incubator temperature 35 C has the largest error which is 2.98 % and the smallest error 0.86%. The conclusion of this study is that the design can work well in measuring the temperature of the incubator, as well as the system for storing readings using the SD card Design and sending data using the internet network can work well.
Eight Channel Temperature Monitoring using Thermocouple Sensors (type K) Based on Internet of Thing using ThinkSpeak Platform Candra Prastyadi; Bedjo utomo; Her Gumiwang Ariswati; Dyah Titisari; Sumber Sumber; A. Senthil Kumar
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 5 No 1 (2023): January
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v5i1.276

Abstract

A laboratory incubator is a device used to incubate a breed. a very important condition in the procedure Incubator is the optimal temperature conditions for microorganisms to grow. The incubator is equipped with a temperature controller so that the temperature can be adjusted according to the breed to be raised. Incubators use an oven like dry heat. The purpose of this study was to test and analyze the accuracy of the thermocouple sensor with incubator media in a laboratory incubator calibrator. The main design method uses the 8 MAX 6675 module, the 8 K type Thermocouple module, Arduino Mega, and SD Card data storage. Temperature measurements were measured with a Type K thermocouple sensor. The thermocouple sensor has 8 channels which function to measure the temperature at each camber point of the incubator. The temperature will be stored on the SD card for data analysis and the data can be processed in graphical form. Benchmarking is done using a temperature data logger. This is done so that the design results are below the standard comparison tool. The measurement results on the module compared to the comparison tool obtained the largest error value, namely 3.98%, namely on channel T6 at 35°C with ordinary incubator media and the smallest error on ordinary incubator media at point T6 at 37°C, which is 0.06 % and at 35 C the temperature of the incubator fan has the largest error of 2.98% and the smallest error of 0.86%. the module can perform well by testing the comparison tool at every point
Analysis of BPM and Temperature Data Retrieval in Smartband Design with Android Applications Using the Multiplexing Method Jarot Hadi Wibowo; Her Gumiwang Ariswati; I Dewa Gede Hari Wisana; Shubhrojit Misra
Jurnal Teknokes Vol 16 No 1 (2023): March
Publisher : Jurusan Teknik Elektromedik, POLTEKKES KEMENKES Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/teknokes.v16i1.516

Abstract

Monitoring and recording of patient medical records is very necessary if there are symptoms of a disease that must be taken quickly so that the patient's condition does not worsen and find out how the pattern of a disease attacks the body's condition, this is so that the patient's history data can be checked by a doctor to make it easier and improve the accuracy of the doctor's diagnosis. In its application, the use of two sensors for smartband requires the use of a multiplexer method so that the data does not collide with each other because of the similarity of the two sensor systems. In this study, the Wemos D1 Mini microcontroller was used to process BPM and temperature data, as well as to process the time of using the multiplexer of the two sensors, then the BPM and temperature values ​​were displayed on the OLED screen. Based on the results of the research that has been done, it can be concluded that the smartband has an error (error value) which is still within the allowable tolerance limit of ±5%. This study has the advantage of a portable device design and is very suitable for daily use to monitor the patient's condition, so that patients are not disturbed by the presence of this smartband. This research has a development that is the use of a larger battery so that its use can last longer, then the use of a temperature sensor that has a medical grade standard.
Design and Build a Ventalitor Tester With PIP and PIF Waveform Displays As Validation (PIP) Sandhi Dhanindra; Her Gumiwang Ariswati; Dyah Titisari; Mansour Asghari
Jurnal Teknokes Vol 16 No 2 (2023): June
Publisher : Jurusan Teknik Elektromedik, POLTEKKES KEMENKES Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/teknokes.v16i2.473

Abstract

PIP (Peak Inspiratory Pressure) is the highest level of pressure exerted into the lungs by the ventilator during inhalation. This PIP parameter is important for monitoring because inappropriate PIP values ​​can lead to fatal errors in patients. PIP should be kept below 20 to 25 cm H2O each time ventilation. This PIP control can be a value as well as a waveform. This waveform display is also used to validate the ventilator output. Checking the output on this ventilator is used using a ventilator tester. The purpose of this research is to get the accuracy and precision of the sensor to display the waveform and PIP value of the ventilator output. The procedure of this research is to use MPX5010GP to detect the pressure value on the ventilator and then display the value and waveform of the PIP. From this study, the results of the measurement of accuracy and precision from the MPX5010 sensor to detect PIP and display a waveform graph are said to be good. This is because the highest error value is ±6.27% at the 15 CmH20 setting. While the value of the largest standard deviation at the 15 CmH20 setting is 0.837 and the greatest uncertainty value at the 15 CmH20 setting is 0.033. Then, the largest correction value is found in the 25 CmH20 setting, which is 1.56. PIF monitoring is carried out to maximize service to patients and maximize ventilator care.